19
Operation and Care Manual
Setup and Basic Usage
Pull Sliding
Sleeve
Drape
(Sectioned for illustration purposes)
Connect the scope to the camera by pulling the sliding sleeve and inserting the
scope. Ensure that the light port is aligned with the bottom of the handpiece and
the green dot on the back of the scope lines up with the coupler bump. Release the
sliding sleeve to secure the scope in place. The proper technique must be utilized
in order to preserve the sterile field through the drape that is attached to each
endoscope.
Turn the power on. The monitor will power up within about fifteen (15)
seconds. Hold the distal tip of the scope approximately 2.0cm from a surface with
fine details such as your palm or surgical drape.
Turn the focus adjustment knob in either direction until the fine-detailed
surface at the distal end of the scope comes into optimal focus on the LCD video
monitor.
See also: § 3.2.2 Setup for Each Procedure
3.1.3
Attaching External Video Devices
Summary of Contents for innerVue II