5061-084-REV-1
LIGHTWAVE 3 & 3+ SPECTROPHOTOMETERS USER’S MANUAL
24
USER INTERFACE
Step 10
Proceed to the Calibration screen by selecting
“Next” using the confirm key.
OR
Return to the previous parameters screen by
selecting “Back” using the escape key.
Step 11
Insert the reference sample then take a reference
measurement using the reference key.
The acquired reference sample baseline will be
applied to all subsequent standard and test sample
measurements.
Step 12
Replace the previous sample with the first standard
sample then take a sample measurement using the
confirm key.
Repeat for all standards samples.
Step 13
Proceed to the measurement screen by selecting
“OK” using the confirm key.
OR
Return to the standards screen by selecting “Back”
using the escape key.