BD / BF / ED / FD / FED (E3.1) 02/2022
page 77/116
Chamber size
BF 56
BF 115
BF 260
BF 720
Interior dimensions
(continued)
Interior volume
l /
cu.ft.
59 /
2.08
114 /
4.03
257 /
9.08
728 /
25.71
Steam space volume
l /
cu.ft.
66 /
2.33
127 /
4.49
279 /
9.85
791 /
27.93
Racks
Quantity of racks (regular)
2
2
2
2
Quantity of racks (max.)
4
5
8
16
Max. load per rack (standard rack)
Kg /
lbs
30 /
66
30 /
66
40 /
88
45 /
99
Max. load per rack
(optional perforated shelf)
Kg /
lbs
35 /
77
35 /
77
35 /
77
35 /
77
Max. load per rack
(optional heavy load rack)
Kg /
lbs
50 /
110
70 /
154
80 /
176
70 /
154
Permitted total load
Kg /
lbs
70 / 154
150 /
330
270 / 595
315 /
694
Weight
Weight (empty)
Kg /
lbs
39 /
86
54 /
119
85 /
187
166 /
366
Temperature data
Temperature
range
from … degrees
above ambient
°C /
°F
7 /
12.6
8 /
14.4
7 /
12.6
10 /
18
up to
°C /
°F
100 /
212
100 /
212
100 /
212
100 /
212
Temperature fluctuation at 37 °C /
98.6 °F
± K
0.1
0.1
0.1
0.1
Temperature uniformity (variation)
at 37 °C /
98.6 °F
± K
0.3
0.3
0.3
0.3
Heating up time to 37 °C /
98.6 °F
minutes
8
8
8
15
Recovery time after door was
opened for 30 sec at 37 °C /
98.6 °F
minutes
3
4
4
4
Electrical data
(model versions BF056-230V, BF115-230V, BF260-230V, BF720-230V)
System of protection acc. to EN
60529
IP
20
20
20
20
Nominal
voltage
(+/-10%)
at 50 Hz power frequency
V
230
230
230
230
at 60 Hz power frequency
V
230
230
230
230
Current type
1N~
1N~
1N~
1N~
Nominal power
kW
0.40
0.40
0.90
1.75
Nominal power with option internal
socket
kW
0.90
0.90
1.40
2.25
Power plug of the power cable
Grounded plug
Chamber fuse (external)
5x20 mm / 250V / time-lag (T)
A
6.3
6.3
8.0
12.5
Overtemperature protective device
class 1
°C
120
120
120
120
Installation category acc. to IEC 61010-1
II
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
2
Different electrical data for BF-UL constructed for the USA and Canada
(model versions BF056UL-120V, BF115UL-120V, BF260UL-120V, BF720UL-240V)
Nominal
voltage
(+/-10%)
at 50 Hz power frequency
V
120
120
120
240
at 60 Hz power frequency
V
120
120
120
240
Current type
1N~
1N~
1N~
2~
Power plug of the power cable
NEMA
5-15P
5-15P
5-15P
6-20P
Nominal power
kW
0.40
0.40
1.00
1.85
Chamber fuse (external)
5x20 mm / 250V / time-lag (T)
A
12.5
12.5
12.5
---
Miniature circuit breaker (internal)
A
---
---
---
16
Summary of Contents for BD 115
Page 86: ...BD BF ED FD FED E3 1 02 2022 page 86 116 16 11 Dimensions size 56...
Page 87: ...BD BF ED FD FED E3 1 02 2022 page 87 116...
Page 88: ...BD BF ED FD FED E3 1 02 2022 page 88 116 16 12 Dimensions size 115...
Page 89: ...BD BF ED FD FED E3 1 02 2022 page 89 116...
Page 90: ...BD BF ED FD FED E3 1 02 2022 page 90 116 16 13 Dimensions size 260...
Page 91: ...BD BF ED FD FED E3 1 02 2022 page 91 116...
Page 92: ...BD BF ED FD FED E3 1 02 2022 page 92 116 16 14 Dimensions size 720...
Page 93: ...BD BF ED FD FED E3 1 02 2022 page 93 116...
Page 94: ...BD BF ED FD FED E3 1 02 2022 page 94 116...
Page 96: ...BD BF ED FD FED E3 1 02 2022 page 96 116...
Page 97: ...BD BF ED FD FED E3 1 02 2022 page 97 116 17 2 EU Declaration of Conformity for BF...
Page 98: ...BD BF ED FD FED E3 1 02 2022 page 98 116...
Page 99: ...BD BF ED FD FED E3 1 02 2022 page 99 116 17 3 EU Declaration of Conformity for ED...
Page 100: ...BD BF ED FD FED E3 1 02 2022 page 100 116...
Page 101: ...BD BF ED FD FED E3 1 02 2022 page 101 116 17 4 EU Declaration of Conformity for FD...
Page 102: ...BD BF ED FD FED E3 1 02 2022 page 102 116...
Page 103: ...BD BF ED FD FED E3 1 02 2022 page 103 116 17 5 EU Declaration of Conformity for FED...
Page 104: ...BD BF ED FD FED E3 1 02 2022 page 104 116...
Page 106: ...BD BF ED FD FED E3 1 02 2022 page 106 116...
Page 107: ...BD BF ED FD FED E3 1 02 2022 page 107 116...
Page 108: ...BD BF ED FD FED E3 1 02 2022 page 108 116...
Page 109: ...BD BF ED FD FED E3 1 02 2022 page 109 116...
Page 117: ......