BD / BF / ED / FD / FED (E3.1) 02/2022
page 79/116
Chamber size
ED 56
ED 115
ED 260
ED 720
Temperature data
Temperature
range
from … degrees
above ambient
°C /
°F
5 /
9
5 /
9
5 /
9
5 /
9
up to
°C /
°F
300 /
572
300 /
572
300 /
572
300 /
572
Temperature fluctuation at 150 °C /
302 °F
± K
0.5
0.4
1.0
0.8
Temperature uniformity (variation)
at 150 °C /
302 °F
± K
2.5
2.0
2.0
3,2
Heating up time to 150 °C /
302 °F
minutes
45
45
55
85
Recovery time after door was
opened for 30 sec at 150 °C /
302 °F
minutes
35
16
20
25
Electrical data
(model versions ED056-230V, ED115-230V, ED260-230V, ED720-400V)
IP system of protection acc. to EN
60529
IP
20
20
20
20
Nominal
voltage
(+/-10%)
at 50 Hz power frequency
V
230
230
230
400
at 60 Hz power frequency
V
230
230
230
400
Current type
1N~
1N~
1N~
3N~
Nominal power
kW
1.05
1.25
2.25
4.10
Power plug of the power cable
Grounded plug
Chamber fuse (external)
5x20 mm / 250V / time-lag (T)
A
6.3
6.3
12.5
---
Miniature circuit breaker (internal)
A
---
---
---
16
Overtemperature protective device
class 1
°C
330
330
330
330
Installation category acc. to IEC 61010-1
II
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
2
Different electrical data for ED-UL constructed for the USA and Canada
(model versions ED056UL-120V, ED115UL-120V, ED260UL-240V)
Nominal
voltage
(+/-10%)
at 50 Hz power frequency
V
120
120
240
---
at 60 Hz power frequency
V
120
120
240
---
Current type
1N~
1N~
2~
---
Power plug of the power cable
NEMA
5-15P
5-15P
6-20P
---
Nominal power
kW
1.15
1.35
2.45
---
Chamber fuse (external)
5x20 mm / 250V / time-lag (T)
A
12.5
12.5
---
---
Miniature circuit breaker
(combination element, external)
A
---
---
16
---
Environment-specific data
Energy consumption at 150 °C /
302 °F
Wh/h
180
250
370
700
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C +/- 3 °C /
71.6 °F +/- 5.4 °F
and a power supply voltage fluctuation of +/-10%. Technical data is
determined in accordance to BINDER Factory Standard Part 2:2015 and DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up times may vary according to the load.
Summary of Contents for BD 115
Page 86: ...BD BF ED FD FED E3 1 02 2022 page 86 116 16 11 Dimensions size 56...
Page 87: ...BD BF ED FD FED E3 1 02 2022 page 87 116...
Page 88: ...BD BF ED FD FED E3 1 02 2022 page 88 116 16 12 Dimensions size 115...
Page 89: ...BD BF ED FD FED E3 1 02 2022 page 89 116...
Page 90: ...BD BF ED FD FED E3 1 02 2022 page 90 116 16 13 Dimensions size 260...
Page 91: ...BD BF ED FD FED E3 1 02 2022 page 91 116...
Page 92: ...BD BF ED FD FED E3 1 02 2022 page 92 116 16 14 Dimensions size 720...
Page 93: ...BD BF ED FD FED E3 1 02 2022 page 93 116...
Page 94: ...BD BF ED FD FED E3 1 02 2022 page 94 116...
Page 96: ...BD BF ED FD FED E3 1 02 2022 page 96 116...
Page 97: ...BD BF ED FD FED E3 1 02 2022 page 97 116 17 2 EU Declaration of Conformity for BF...
Page 98: ...BD BF ED FD FED E3 1 02 2022 page 98 116...
Page 99: ...BD BF ED FD FED E3 1 02 2022 page 99 116 17 3 EU Declaration of Conformity for ED...
Page 100: ...BD BF ED FD FED E3 1 02 2022 page 100 116...
Page 101: ...BD BF ED FD FED E3 1 02 2022 page 101 116 17 4 EU Declaration of Conformity for FD...
Page 102: ...BD BF ED FD FED E3 1 02 2022 page 102 116...
Page 103: ...BD BF ED FD FED E3 1 02 2022 page 103 116 17 5 EU Declaration of Conformity for FED...
Page 104: ...BD BF ED FD FED E3 1 02 2022 page 104 116...
Page 106: ...BD BF ED FD FED E3 1 02 2022 page 106 116...
Page 107: ...BD BF ED FD FED E3 1 02 2022 page 107 116...
Page 108: ...BD BF ED FD FED E3 1 02 2022 page 108 116...
Page 109: ...BD BF ED FD FED E3 1 02 2022 page 109 116...
Page 117: ......