C 150 (E2) 06/2011
page 62/78
CAUTION
The glass door and inner chamber become hot during sterilization.
Danger of burning.
∅
Do NOT touch the glass door and inner surfaces during sterilization.
CAUTION
Interruption of the temperature reaction time.
Ineffective sterilization.
∅
Do NOT open the unit doors during sterilization.
•
The sterilization cycle ends after approx. 12 hours, when the lower display shows “
End
”.
Figure 35: Alternating display at the end of sterilization
•
Turn off the unit or press “
180 °C
”, or open the outer door.
•
When the inner chamber has cooled down to a value below 60 °C /
140°F
, turn off the unit and plug in
the CO
2
sensor.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C /
140°F
.
•
Turn on the unit (chap. 5).
The unit is now ready to operate.
15.5 Aborting the hot-air sterilization
If the sterilization procedure is aborted prematurely, whether effective sterilization has occurred depends
on the time that has elapsed:
•
Aborting sterilization after less than 6 hours: Prevents effective sterilization.
•
Aborting sterilization after more than 6 hours: The unit is definitely in the cooling-down phase,
meaning that the necessary duration for the proper sterilization phase has occurred.
Aborting sterilization during the cooling-down phase (after more than 6 hours)
The duration of the entire sterilization is approx. 12 hours. If you want to shorten the sterilization proce-
dure in order to save time, you can abort it during the cooling-down phase, i.e. no sooner than after 6
hours. At this point, the inner temperature is still approx. 140 °C
/ 284°F.