FED / FED-UL (E2) 01/2022
page 48/60
Electrical data FED400-400V
IP system of protection acc. to EN 60529
20
Nominal voltage
(+/-10 %)
at 50 Hz power frequency
V
400
at 60 Hz power frequency
V
400
Current type
3N~
Nominal power
kW
3.40
Over-current release category B
3 x 16A internal
Power plug
CEE plug 5 poles
Installation category acc. to IEC 61010-1
II
Pollution degree acc. to IEC 61010-1
2
Different electrical data for FED-UL constructed for the USA and Canada (FED400UL-208V)
Nominal voltage (±10 %) at 60 Hz power frequency
V
208
Current type
3N~
Power plug
NEMA
L21-20P
Chamber fuse 6,3 x 32 mm
250V / super-time-lag TT
A
16
3 x internal
Environment-specific data
Noise level (mean value)
dB (A)
< 55
Energy consumption
at 70 °C /
158 °F
Wh/h
520
at 150 °C /
302 °F
Wh/h
1200
at 300 °C /
572 °F
Wh/h
2340
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C +/- 3 °C /
71.6 °F +/- 5.4 °F
and a power supply voltage fluctuation of +/- 10. Technical data is
determined in accordance to BINDER Factory Standard Part 1:2015 following DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up times may vary according to the load.
12.5 Equipment and options (extract)
To operate the chamber, use only original BINDER accessories or accessories / components
from third-party suppliers authorized by BINDER. The user is responsible for any risk arising
from using unauthorized accessories.
Standard equipment
Microprocessor temperature controller with LED display and several time functions
Controller Timer functions: Delayed ON, delayed Off and temperature dependent delayed OFF
Temperature safety device class 2 acc. to DIN 12880:2007 with visual temperature alarm
Adjustable ramp function
Rear exhaust duct, internal diameter
50
mm /
1,97 inch
with ventilation slide
Adjustable air change by means of rear exhaust duct (50 mm) with ventilation flap and front ventilation
slide
2 racks, chrome-plated
RS 422 interface for APT-COM™ 4 Multi Management Software, or switch over to printer output with RS
232/RS 422 interface converter