12.4 FED / FED-UL technical data
Exterior dimensions
Width, net
mm /
inch
1234 /
48.58
Height, gross (incl. feet/castors)
mm /
inch
1022 /
40.24
Depth
mm /
inch
765 /
30.12
Depth gross (incl. door handle, and exhaust duct)
mm /
inch
855 /
33.66
Wall clearance rear (minimum)
mm /
inch
100 /
3.94
Wall clearance side (minimum)
mm /
inch
160 /
6.30
Exhaust duct, outer diameter
mm /
inch
52 /
2.05
Doors
Number of doors
2
Interior dimensions
Width
mm /
inch
1000 /
39.37
Height
mm /
inch
800 /
31.50
Depth
mm /
inch
510 /
20.08
Interior volume
l /
cu.ft.
400 /
14.3
Steam space volume
l /
cu.ft.
498 /
17.60
Racks
Quantity of racks (regular)
2
Quantity of racks (max.)
10
Maximum load per rack
Kg /
lbs
35 /
77
Permitted total load
Kg /
lbs
90 /
199
Weight
Weight (empty)
Kg /
lbs
145 /
320
Temperature data
Temperature range, 5 °C /
9 °F
above ambient up to
°C /
°F
300 /
572
Temperature fluctuation
at 70 °C /
158 °F
± K
0.1
at 150 °C /
302 °F
± K
0,7
at 300 °C /
572 °F
± K
1.4
Temperature uniformity
(variation)
at 70 °C /
158 °F
± K
0.8
at 150 °C /
302 °F
± K
3,8
at 300 °C /
572 °F
± K
11
Heating up time
to 70 °C /
158 °F
min
15
to 150 °C /
302 °F
min
29
to 300 °C /
572 °F
min
86
Recovery time after door
was opened for 30 sec
at 70 °C /
158 °F
min
5
at 150 °C /
302 °F
min
6
at 300 °C /
572 °F
min
15
Ventilation data
Air change
at 70 °C /
158 °F
x/h
17
at 150 °C /
302 °F
x/h
18
at 300 °C /
572 °F
x/h
16