FED / FED-UL (E2) 01/2022
page 41/60
Fault description
Possible cause
Required measures
Temperature
(continued)
Chamber without function, only
the green “stand-by” LED is lit
Chamber in stand-by mode
Press down the ON/OFF button (5)
until the display lights up.
Chamber without function. Red
alarm pilot lamp of safety device
(7a) is lit.
Safety device class 2 has turned
off the chamber.
Let cool down the chamber and
press down RESET button. Check
the settings of the temperature set-
point and of the safety device class
2 (chap. 7.1). If appropriate, select
suitable limit value.
Safety device class 2 defective. Contact BINDER service.
Temperature inside the
chamber too high, Red alarm
pilot lamp of safety device (7a)
is lit.
Safety device class 3.1 (option)
has responded.
Check the settings of the
temperature set-point and of the
safety device class 3.1 (chap. 7.2).
Chamber without any function.
No power supply.
Check connection to power supply.
Chamber fuse has responded.
Check chamber fuse and replace it
if appropriate. If it responds again,
contact BINDER service.
Controller defective.
Contact BINDER service.
Deviations from the indicated
heating-up times.
Chamber fully loaded.
Charge the chamber less or
consider longer heating-up times.
Controller
Message „1999” in the controller
display
Sensor rupture between sensor
and controller.
Contact BINDER service.
The controller returns to Normal
Display from any level.
No button was hit for more than
approx. 30 sec.
Repeat entries, enter the values
rapidly.
10.4 Sending the chamber back to BINDER GmbH
If you send a BINDER product to us for repair or any other reason, we will only accept the product upon
presentation of an authorization number that has previously been issued to you. We will issue an
authorization number
(RMA number) after receiving your complaint either in writing or by telephone
prior
to your sending the BINDER product back to us. The authorization number will be issued following receipt
of the information below:
•
BINDER product type and serial number
•
Date of purchase
•
Name and address of the dealer from which you bought the BINDER product
•
Exact description of the defect or fault
•
Complete address contact person and availability of that person
•
Exact location of the BINDER product in your facility
•
A contamination clearance certificate (chap. 14) must be faxed in advance
The authorization number must be applied to the packaging in such a way that it can be easily recognized
or be recorded clearly in the delivery documents.
For security reasons we cannot accept a chamber delivery if it does not carry an
authorization number.
Return address:
BINDER GmbH
Abteilung Service
Gänsäcker 16
78502 Tuttlingen
Germany
Summary of Contents for 9010-0216
Page 52: ...FED FED UL E2 01 2022 page 52 60...
Page 54: ...FED FED UL E2 01 2022 page 54 60...
Page 61: ......