FED / FED-UL (E2) 01/2022
page 39/60
10. Maintenance and service, troubleshooting, repair, testing
10.1 General information, personnel qualification
•
Maintenance
See chap. 10.2
•
Simple troubleshooting
Chap. 10.3 describes troubleshooting by operating personnel. It does not require technical intervention
into the chamber, nor disassembly of chamber parts.
For personnel requirements please refer to chap. 1.1.
•
Detailed troubleshooting
If errors cannot be identified with simple troubleshooting, further troubleshooting must be performed by
BINDER Service or by BINDER qualified service partners or technicians, in accordance with the
description in the Service Manual.
For personnel requirements please refer to the Service Manual.
•
Repair
Repair of the chamber can be performed by BINDER Service or by BINDER qualified service partners
or technicians, in accordance with the description in the Service Manual.
After maintenance, the chamber must be tested prior to resuming operation.
•
Electrical testing
To prevent the risk of electrical shock from the electrical equipment of the chamber, an annual repeat
inspection as well as a test prior to initial startup and prior to resuming operation after maintenance or
repair, are required. This test must meet the requirements of the competent public authorities. We
recommend testing under DIN VDE 0701-0702:2008 in accordance with the details in the Service
Manual.
For personnel requirements please refer to the Service Manual.
10.2 Maintenance intervals, service
DANGER
Electrical hazard during live maintenance work.
Deadly electric shock.
∅
The chamber must NOT become wet during operation or maintenance work.
∅
Do NOT remove the rear panel of the chamber.
Before conducting maintenance work, turn off the chamber at the main power switch
and disconnect the power plug.
Ensure all maintenance work is conducted by licensed electricians or experts
authorized by BINDER.
Ensure regular maintenance work is performed at least once a year and that the legal requirements are
met regarding the qualifications of service personnel, scope of testing and documentation.
The warranty becomes void if maintenance work is conducted by non-authorized personnel.
Replace the door gasket only when cold. Otherwise, the door gasket may become damaged.