Binder 9010-0216 Operating Manual Download Page 1

 

Issue 01/2022 

Art. No. 7001-0070

 

 
 
 
 
 
 

Operating Manual

 

Translation of the original operating manual 

 
 
 
 
 

FED / FED-UL (E2) 

 

Drying and heating ovens  

with forced convection and enhanced timer 

functions

 

with microprocessor temperature controller 

 

Model 

Model version 

Art. No. 

FED 400 (E2) 

FED400-400V 

9010-0216, 9110-0216 

FED 400-UL (E2) 

FED400UL-208V 

9010-0217, 9110-0217 

 
 
 
 

BINDER GmbH 

 Address: Post office box 102, 78502 Tuttlingen, Germany 

 Phone: +49 7462 2005 0 

 Fax: +49 7462 2005 100 

 Internet: http://www.binder-world.com 

 E-mail: [email protected] 

 Service Hotline: +49 7462 2005 555 

 Service Fax: +49 7462 2005 93 555 

 Service E-Mail: [email protected] 

 Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3 

 Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03 

 Service Hotline Russia and CIS: +7 495 988 15 16 

 

Summary of Contents for 9010-0216

Page 1: ...FED 400 UL E2 FED400UL 208V 9010 0217 9110 0217 BINDER GmbH Address Post office box 102 78502 Tuttlingen Germany Phone 49 7462 2005 0 Fax 49 7462 2005 100 Internet http www binder world com E mail inf...

Page 2: ...ng and transportation 17 3 3 Storage 17 3 4 Location of installation and ambient conditions 18 4 INSTALLATION 19 4 1 Electrical connection 19 4 2 Connection to a suction plant optional 20 5 START UP 2...

Page 3: ...he transport packing 42 11 2 Decommissioning 42 11 3 Disposal of the chamber in the Federal Republic of Germany 42 11 4 Disposal of the chamber in the member states of the EU except for the Federal Re...

Page 4: ...nificant risks DANGER Dangers due to failure to observe the instructions and safety precautions Serious injuries and chamber damage Risk of death Observe the safety instructions in this Operating Manu...

Page 5: ...communities of states This includes word marks position marks word figurative marks design configurations figurative marks and design patents Patent Information BINDER products categories of products...

Page 6: ...sive atmosphere Stability hazard Lifting hazard Suffocation hazard Harmful substances Risk of corrosion and or chemical burns Biohazard Pollution Hazard Mandatory action signs Mandatory regulation Rea...

Page 7: ...that could result in direct or indirect injury or property damage 1 5 Localization position of safety labels on the chamber The following labels are located on the chamber Pictograms Warning signs Se...

Page 8: ...l voltage 10 at the indicated power frequency 400 V 60 Hz 3 N Current type Symbols on the type plate Symbol Valid for Information All chambers CE conformity marking All chambers Electrical and electro...

Page 9: ...arts To operate the chamber use only original BINDER accessories or accessories from third party suppliers authorized by BINDER The user is responsible for any risk caused by using unauthorized access...

Page 10: ...adequate measures to exclude such risks prior to putting the chamber into operation DANGER Electrical hazard by water entering the chamber Deadly electric shock The chamber must NOT become wet during...

Page 11: ...achine components Such ingredients include in particular acids and halides Any corrosive damage caused by such ingredients is excluded from liability by BINDER GmbH The chamber does not dispose of any...

Page 12: ...s than those described in chap 1 8 are not approved This expressly includes the following misuses the list is not exhaustive which pose risks despite the inherently safe construction and existing tech...

Page 13: ...uired to prevent them are listed in the Operating Manual Moreover the operator must take measures to minimize hazards from unavoidable residual risks This includes in particular issuing operating inst...

Page 14: ...ing instructions Keep these operating instructions with the chamber at all times in a place where they are clearly visible They must be comprehensible and written in the language of the employees 1 12...

Page 15: ...electrically and are ventilated by fan assisted forced air circulation They FED are equipped with a temperature safety device according to DIN12880 chap 7 The APT line preheating chamber system guara...

Page 16: ...Inform the carrier immediately if transportation damage has occurred The final tests of the manufacturer may have caused traces of the racks on the inner surfaces This has no impact on the function a...

Page 17: ...with transport straps for transport Do NOT lift or transport the chamber using the door handle or the door Do NOT lift chambers by hand Place chambers using technical devices fork lifter on the trans...

Page 18: ...or air solvent mixtures AWAY from the vicinity of the chamber Ambient conditions Permissible ambient temperature range during operation 18 C up to 40 C At elevated ambient temperature values fluctuat...

Page 19: ...UL Listed Power supply cord UL category ELBZ SJT 3x14 AWG 2 08 mm C13L For outside USA use a certified power supply cord according to national requirements Prior to connection and start up check the p...

Page 20: ...o reduce odors quickly we recommend heating up the chamber to its nominal temperature for one day and in a well ventilated location 1 Insert the power plug into a suitable socket chap 4 1 2 Turn on ch...

Page 21: ...l value display The actual temperature value example 22 C is displayed 1 Press button The display shows alternately SP and the previous temperature set point example 60 C 2 With the buttons enter a se...

Page 22: ...chamber is operating the controller is in normal display actual value display The actual temperature value example 22 C is displayed 1 Press button The display shows alternately SPr and the actual tem...

Page 23: ...on There are two possible time functions Continuous operation The display shows alternately t1 time function and the time function Continuous operation t inf The heating is permanently active independ...

Page 24: ...onds Timer operation Timer not programmed or run down Press button to return to normal display actual value display automatically after approx 30 sec 6 3 2 Continuous operation 1 Press the time manage...

Page 25: ...The use of this time depends on the timer function selected in the user menu chap 6 4 4 4 Press button to return to normal display actual value display automatically after approx 30 seconds The actua...

Page 26: ...utton Timer function chap 6 4 4 Press button Setting the interface mode and if appropriate the printer interval chap 6 4 5 Figure 5 Press down button for approx 5 seconds Press button to return to nor...

Page 27: ...sed if required The use of temperature ramps may result in the heating up times being considerably slowed down The entry in C min or in F min meaning the nominal value gradient and limits the maximum...

Page 28: ...gradient from the previously entered set point to the new one SP The actual value follows the set point value About set point display during ramp operation see chap 6 2 6 4 3 Chamber addressing If sev...

Page 29: ...low the set point After the defined time has expired the heating is turned off Delayed on setting 2 After the time set has passed the heating is turned on and remain in continuous operation 1 Press do...

Page 30: ...timer function 5 Press again button The display alternately shows PFCt and the actual setting of the interface mode 6 Set the desired interface mode with buttons Settings Modbus 0 printer 1 In case of...

Page 31: ...n normal display press the time management button The controller displays the actual time function If necessary select the time function Timer operation chap 6 3 1 In the entry level enter the desired...

Page 32: ...the safety device by pressing the reset button 7b and turn on the chamber as described Figure 6 Temperature safety device class 2 Function The safety device class 2 is functionally and electrically i...

Page 33: ...changes of the set point Function check Check the temperature safety device class 2 at appropriate intervals for its functionality It is recommended that the authorized operating personnel should per...

Page 34: ...ated chap 7 3 the buzzer sounds as an additional signal You can turn it off with switch 11 4 The optimum setting of the safety device is obtained by turning the control knob clockwise by approximately...

Page 35: ...are to be recorded via a PC each one must be allocated a unique address Addressing is performed via the chamber controller see chap 6 4 3 8 2 HEPA fresh air filter option With this option the introdu...

Page 36: ...l Wipe the surfaces with a moistened towel In addition you can use the following cleaning agents Exterior surfaces inner chamber racks door gaskets Standard commercial cleaning detergents free from ac...

Page 37: ...he neutral cleaning agent Recommended precautions To protect the eyes use sealed protective goggles Wear gloves Suitable protective gloves in full contact with media butyl or nitrile rubber penetratio...

Page 38: ...inutes All inflammable goods must be removed from the interior before 2 Spray the inner chamber with an appropriate disinfectant Before start up the chamber must be absolute dry and ventilated because...

Page 39: ...rom the electrical equipment of the chamber an annual repeat inspection as well as a test prior to initial startup and prior to resuming operation after maintenance or repair are required This test mu...

Page 40: ...epair Repaired chambers must comply with the BINDER quality standards Fault description Possible cause Required measures General Chamber without function No power supply Check connection to power supp...

Page 41: ...Message 1999 in the controller display Sensor rupture between sensor and controller Contact BINDER service The controller returns to Normal Display from any level No button was hit for more than appro...

Page 42: ...m the power supply pull the power plug When turning off the main power switch ON OFF 10 the stored parameters remain saved Temporal decommissioning See indications for appropriate storage chap 3 3 Fin...

Page 43: ...ecycling according to Directive 2012 19 EU The devices must be free from toxic infectious or radioactive substances in order to eliminate any health hazards to the employees of the recycling companies...

Page 44: ...posed of professionally at a recycling company which is certified according to conversion of the Directive 2012 19 EU into national law or Instruct the distributor who sold you the device to dispose o...

Page 45: ...ection from the chamber A chamber from which all toxic substances or sources of infection cannot be safely removed must be considered as special waste according to national law Dispose of it according...

Page 46: ...olume illustrated below is calculated as follows c c b a C B A b a A B C Internal dimensions W H D a b c Wall clearances a 0 1 x A b 0 1x B c 0 1 x C VUSE A 2a x B 2b x C 2c Figure 9 Determination of...

Page 47: ...eam space volume l cu ft 498 17 60 Racks Quantity of racks regular 2 Quantity of racks max 10 Maximum load per rack Kg lbs 35 77 Permitted total load Kg lbs 90 199 Weight Weight empty Kg lbs 145 320 T...

Page 48: ...ce to BINDER Factory Standard Part 1 2015 following DIN 12880 2007 All indications are average values typical for chambers produced in series We reserve the right to change technical specifications at...

Page 49: ...sconnectable audible over temperature alarm Door with window and interior lightning Lockable door FKM door gasket temperature resistant up to 200 C HEPA Fresh air filter class H 14 DIN EN 1822 Measure...

Page 50: ...e class 1 5006 0037 Rack lockings 4 pieces 8012 0531 HEPA Fresh air filter class H 14 DIN EN 1822 2009 8012 0076 Neutral cleaning agent 1 kg 1002 0016 For information on components not listed here ple...

Page 51: ...FED FED UL E2 01 2022 page 51 60 13 Certificates and declarations of conformity 13 1 EU Declaration of Conformity...

Page 52: ...FED FED UL E2 01 2022 page 52 60...

Page 53: ...FED FED UL E2 01 2022 page 53 60 13 2 Certificate for the GS mark of conformity of the Deutsche Gesetzliche Unfallversicherung e V German Social Accident Insurance DGUV...

Page 54: ...FED FED UL E2 01 2022 page 54 60...

Page 55: ...bevor das Ger t Bauteil eintrifft Eine weitere Kopie soll dem Ger t Bauteil beigef gt sein Ggf ist auch die Spedition zu informieren Incomplete information or non conformity with this procedure will i...

Page 56: ...dous substances weder giftige noch sonstige gef hrliche Stoffe enth lt oder solche anhaften That eventually generated reaction products are non toxic and also do not represent a hazard auch evtl entst...

Page 57: ...gaben entsteht haften und BINDER gegen eventuell entstehende Schadenanspr che Dritter freistellen We are aware that in accordance with Article 823 of the German Civil Code BGB we are directly liable w...

Page 58: ...t any time Take notice of shipping laws and regulations Please fill Reason for return request Duplicate order Duplicate shipment Demo Page one completed by sales Power Plug Voltage 115V 230 V 208 V 24...

Page 59: ..._________________________________________________________________ c ____________________________________________________________________________ 3 2 Safety measures required for handling the list unde...

Page 60: ...for all damages that are a consequence of incomplete or incorrect information provided by us and that we will indemnify and hold harmless BINDER Inc from eventual damage claims by third parties Name _...

Page 61: ......