0, 1, 2, 4
No. of divisions for
stability signal
Defines the number of divisions by which the
weight may vary while the weight stable signal
is maintained. On terminals subject to metrical
verification, this parameter must be set to 0
(zero) or 1.
high resolution
YES, NO
High resolution operating parameter. This
function is useful for weighing tests and scale
calibration, as it allows the weight reading to
be displayed in tenths of a division. It is not
enabled during normal terminal operation. It is
used primarily during scale testing.
zero track
0e, 0.25e, 0.5e, 1e
Defines the range around zero within which
zero tracking is performed to small gradual
variations in the weight reading, caused, for
example, by dust settling on the scale.
✔
For dosing systems, set the parameter to 0
(zero)
✔
When zero tracking is disabled by selecting
0 (zero), the forced zero function is also
automatically disabled. This applies even if
the forced zero parameter is set to a value
other than 0 (zero).
D400
3-27