Replace 1 cell
It may be necessary to replace a cell in the
system. In this case, the angular coefficient
can be calculated for the replaced cell without
having to calibrate all the other cells.
Proceed in the following way:
unload the scale and press
READY
.
✔
The cells appear on the display arranged in
sequence and without identification
numbers. This display remains for the time
required to acquire the minimum dead load.
✔
Load the sample weight on the replaced
cell when requested by the terminal and
press
READY
to acquire its value;
✔
load the sample weight on the cell
diametrically opposite the replaced one
when requested by the terminal and press
READY
.
The terminal will calculate the angular
coefficient for the replaced cell.
✔
Press
SAVE
to record the angular
coefficient in the terminal and cell.
There terminal will indicate if the angular
coefficient value is beyond the tolerated limits.
In this case, press
REPEAT
to repeat the
calculations.
✔
Switch off the terminal.