background image

Safety advice before use

  Before use please read and understand this 

manual and familiarise yourself with the 
layout of the controls – see figure 1.

  If the mains supply is interrupted the unit will 

restart with its current settings on the 
restoration of the electricity supply.

  Switch off when not in use.

  Position the instrument on a firm level surface 

away from any heat sensitive or flammable 
material.

  Do not use the instrument with flammable 

liquids or in a hazardous atmosphere.

  Never move or carry the instrument until it 

has been switched off and allowed to cool.

  In either bath, never mix oil and water.

  Always ensure that a suitable heating 

medium is used in the RE3000OB.

RE300DB

Fill the Teflon coated bowl to the required level 
with cold water. To avoid scale build up and the 
necessity for cleaning it is recommended that the 
bowl be filled with distilled or deionised water.

The maximum capacity of the bath is 3 litres and 
it should not be filled to within less than 2cm 
from the top of the bowl.

During use water will be lost from the bath by 
evaporation. For maximum efficiency water 
should be added periodically and the level 
maintained.

Do not fill the RE300DB bath with any liquid 
other than water.

RE300OB

The RE300OB is intended for use with either oil 
or water. 

NOTE: Oil and water should never 

be mixed in the bath.

 Fill the Teflon coated 

bowl to the required level with either oil or 
water, the maximum capacity of the bath is 6.2L 
the fill level is 2cm from the top of the bowl.

If being used with water, during use water will 
be lost from the bath by evaporation. For 
maximum efficiency water should be added 
periodically and the level maintained.

The RE300OB is equipped with a height 
adjustable foot. This can be turned clockwise to 
decrease the height and anti-clockwise to 
increase it. Turn the adjuster by hand whilst 
clamping the locking nut by hand or use a 
22mm wrench, see Fig 3.

3

Adjuster

Lock nut

To tighten

To loosen

Fig 3.

Summary of Contents for Stuart RE300DB

Page 1: ...Version 1 0 Rotary Evaporator Bath Instructions for use RE300DB RE300OB C heat digital water bath RE300DB set...

Page 2: ...heat C set oil bath RE300OB Fig 1 Operation panel Power switch Bath base Bath cover Bath Reset button for over temperature protector Fused IEC Connector RE300OB...

Page 3: ...C heat heat C set oil bath RE300OB digital water bath RE300DB set Fig 2 Control panel Power switch Bath cover Main unit case Bath Reset button for over temperature protector Fused IEC Connector RE300D...

Page 4: ...ied with two mains leads fitted with IEC plugs for connection to the instrument One has a U K 3 pin plug and the other has a 2 pin Shuko plug for connection to the mains Choose the lead appropriate fo...

Page 5: ...ed water The maximum capacity of the bath is 3 litres and it should not be filled to within less than 2cm from the top of the bowl During use water will be lost from the bath by evaporation For maximu...

Page 6: ...heat light will turn off Cleaning Where used with water periodically de scale the bath using a domestic kettle descaler suitable for metal kettles Follow the instructions for use supplied with the cle...

Page 7: ...ee 3 years In the event of a justified claim Bibby Scientific Ltd will replace any defective component or replace the unit free of charge This warranty does NOT apply if damage is caused by fire accid...

Page 8: ...the terminal of the local service representative heater is off Temperature controller is impaired Upper temperature limit alarm Heater can not be turned off Stop operation and consult a works because...

Page 9: ...Notes...

Page 10: ...ications We cannot be sure that other equipment used in its vicinity will meet these standards and so we cannot guarantee that interference will not occur in practice Where there is a possibility that...

Page 11: ......

Page 12: ...edex France Tel 33 1 64 45 13 13 Fax 33 1 64 45 13 00 e mail bsf bibby scientific fr www bibby scientific com Bibby Scientific Italia Srl Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milan...

Reviews: