Before use
Thank you for purchasing this Stuart
product. To get the best performance from
the equipment, and for your own safety,
please read these instructions carefully
before use.
If the equipment is not used in the manner
described in this manual the protection
provided by the equipment may be
impaired.
This equipment is designed to operate under the
following conditions:-
❖
For indoor use only
❖
Use in a well ventilated area
❖
Ambient temperature range +5°C to +35°C
❖
Altitude to 2000m
❖
Relative humidity not exceeding 85%
❖
Mains supply fluctuation not exceeding 10%
❖
Overvoltage category II IEC60364-4-443
❖
Pollution degree 2 IEC664
Electrical installation
Before connection please read and
understand this manual and ensure that
the line supply is suitable. This water bath
requires a supply rated at 230V, 50 or
60Hz, single phase, ~.
Power consumption is:
RE300DB - 1.1kVA
RE300OB - 1.4kVA
THIS INSTRUMENT MUST BE EARTHED
The unit is supplied with two mains leads fitted
with IEC plugs for connection to the instrument.
One has a U.K. 3 pin plug and the other has a 2
pin “Shuko” plug for connection to the mains.
Choose the lead appropriate for your electrical
installation and discard the other.
Should neither lead be suitable, take the lead
with the U.K. plug and replace the plug with a
suitable alternative. This involves cutting off the
moulded plug, preparing the cable and
connecting to the rewireable plug in accordance
with its instructions.
IT IS IMPORTANT THAT THIS OPERATION
SHOULD ONLY BE UNDERTAKEN BY A
QUALIFIED ELECTRICIAN
The wires in the mains cable are coloured as
follows:
BROWN
- LIVE
BLUE
- NEUTRAL
GREEN/YELLOW - EARTH
The unit is fitted with an IEC socket at the rear
of the instrument for connection of the mains
lead. The appropriate mains lead should be
connected BEFORE connection to the mains
supply.
Should the mains lead need replacement a cable
of 1mm
2
of harmonised code H05W-F connected
to an IEC 320 plug should be used.
N.B. The UK mains lead is protected by a 10A
fuse mounted in the plug top.
IF IN DOUBT CONSULT A QUALIFIED
ELECTRICIAN