Bibby Sterilin Stuart RE300DB Instructions For Use Manual Download Page 8

Problem 

Cause 

Solution

Display does not work even  

No mains power 

Check mains power supply

if the power is switched on 

 

Fuse has blown 

Replace the fuse, if fuse blows

  

 

again immediately stop  

 

 

 

operation and consult a local    

 

 

service representative.

 

Bath overheat protection  

Turn off the power switch and

 

has tripped either because  

allow the unit to cool. Press the

 

bath was heated without  

reset switch on the back of the

 

oil/water or external  

unit, see Fig 1 & 2. Turn the

 

temperature exceeds 35°C 

power back on. If overheat trips  

 

 

again consult a local service    

 

 

representative.

Temperature does not rise  

Temperature has not been set 

Change the set temperature

(Heat light does not illuminate) 

 

 

Set temperature is below 

 

current bath temperature

 

Temperature does not rise  

Heater is disconnected, 

Stop operation and consult a    

(Heat light illuminates) 

or the terminal of the  

local service representative

 

heater is off 

 

 

 

Temperature controller is 

 

impaired

 

Upper temperature limit alarm  

Heater can not be turned off 

Stop operation and consult a 

works. 

because of malfunction of over  local service representative

Display shows "FF" 

heat protection and measured

 

value exceeds maximum 

 

temperature.

 

Lower temperature limit alarm 

Ambient temperature is lower   Adjust the ambient 

works.  

than 0°C 

temperature to between 

Display shows "--" 

 

5°C and 35°C.

 

Measured temperature is 

 

lower than 0°C

 

Display flashes "FF" 

Temperature sensor shorted 

Stop operation and consult a    

 

 

local service representative

Display flashes "--" 

Temperature sensor id 

 

disconnected or the terminal 

 

of the sensor is off

 

During controlled heating the  

Temperature detection range 

Press and hold the up and down

heat lamp turns off for  

is too wide. 

keys simultaneously until the

prolonged periods of time 

  

display shows "dF". Press the up  

 

 

and down keys to set the  

 

 

 

detection range to 0.5. Press the  

 

 

set key to return to normal  

 

 

 

operation.

Fault finding

6

Summary of Contents for Stuart RE300DB

Page 1: ...Version 1 0 Rotary Evaporator Bath Instructions for use RE300DB RE300OB C heat digital water bath RE300DB set...

Page 2: ...heat C set oil bath RE300OB Fig 1 Operation panel Power switch Bath base Bath cover Bath Reset button for over temperature protector Fused IEC Connector RE300OB...

Page 3: ...C heat heat C set oil bath RE300OB digital water bath RE300DB set Fig 2 Control panel Power switch Bath cover Main unit case Bath Reset button for over temperature protector Fused IEC Connector RE300D...

Page 4: ...ied with two mains leads fitted with IEC plugs for connection to the instrument One has a U K 3 pin plug and the other has a 2 pin Shuko plug for connection to the mains Choose the lead appropriate fo...

Page 5: ...ed water The maximum capacity of the bath is 3 litres and it should not be filled to within less than 2cm from the top of the bowl During use water will be lost from the bath by evaporation For maximu...

Page 6: ...heat light will turn off Cleaning Where used with water periodically de scale the bath using a domestic kettle descaler suitable for metal kettles Follow the instructions for use supplied with the cle...

Page 7: ...ee 3 years In the event of a justified claim Bibby Scientific Ltd will replace any defective component or replace the unit free of charge This warranty does NOT apply if damage is caused by fire accid...

Page 8: ...the terminal of the local service representative heater is off Temperature controller is impaired Upper temperature limit alarm Heater can not be turned off Stop operation and consult a works because...

Page 9: ...Notes...

Page 10: ...ications We cannot be sure that other equipment used in its vicinity will meet these standards and so we cannot guarantee that interference will not occur in practice Where there is a possibility that...

Page 11: ......

Page 12: ...edex France Tel 33 1 64 45 13 13 Fax 33 1 64 45 13 00 e mail bsf bibby scientific fr www bibby scientific com Bibby Scientific Italia Srl Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milan...

Reviews: