Problem
Cause
Solution
Display does not work even
No mains power
Check mains power supply
if the power is switched on
Fuse has blown
Replace the fuse, if fuse blows
again immediately stop
operation and consult a local
service representative.
Bath overheat protection
Turn off the power switch and
has tripped either because
allow the unit to cool. Press the
bath was heated without
reset switch on the back of the
oil/water or external
unit, see Fig 1 & 2. Turn the
temperature exceeds 35°C
power back on. If overheat trips
again consult a local service
representative.
Temperature does not rise
Temperature has not been set
Change the set temperature
(Heat light does not illuminate)
Set temperature is below
current bath temperature
Temperature does not rise
Heater is disconnected,
Stop operation and consult a
(Heat light illuminates)
or the terminal of the
local service representative
heater is off
Temperature controller is
impaired
Upper temperature limit alarm
Heater can not be turned off
Stop operation and consult a
works.
because of malfunction of over local service representative
Display shows "FF"
heat protection and measured
value exceeds maximum
temperature.
Lower temperature limit alarm
Ambient temperature is lower Adjust the ambient
works.
than 0°C
temperature to between
Display shows "--"
5°C and 35°C.
Measured temperature is
lower than 0°C
Display flashes "FF"
Temperature sensor shorted
Stop operation and consult a
local service representative
Display flashes "--"
Temperature sensor id
disconnected or the terminal
of the sensor is off
During controlled heating the
Temperature detection range
Press and hold the up and down
heat lamp turns off for
is too wide.
keys simultaneously until the
prolonged periods of time
display shows "dF". Press the up
and down keys to set the
detection range to 0.5. Press the
set key to return to normal
operation.
Fault finding
6