40
Voltage dips,
short interrup-
tions and
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 V/m
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the ME 90 , inclu-
ding cables, than the recommended separation distance
calculated from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
Recommended separation distance:
d
= 1.2
d
= 1.2
80 MHz to 800 MHz
d
= 2.3
800 MHz to 2.5 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,
a
should be less than
the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment mar-
ked with the following symbol:
Power fre-
quency (50/60
Hz) magnetic
field IEC 61000-
4-8
3 A/m
Not applicable
Not applicable
NOTE
UT is the a.c. mains voltage prior to application of the test level