c
7 "Channel Overview" Menu
b
Rev. 04, 03/2015
91
Adding and Changing Calibration Points
(Continued)
1
Display of calibration mode
2
Read-in value input field
3
Cancel read-in button
Fig. 57
System Menu, Settings - Calibration (reading in calibration point)
The EVU is in "Calibration (CAL)" mode (Display in Fig. 57, Pos. 1), as long
as calibration is being carried out.
The calibration point is automatically created and the current calibration
curve displayed.
Note
Carry out the calibration of the mould level in the same way for
100%.
Note
Ensure that the calibration curve is strictly monotonically increasing.
If not, correct the points or delete them by clicking on the
<Del>
but-
ton (Fig. 55, Pos. 3).
Up to 20 calibration value pairs can be entered.