7 "Channel Overview" Menu
b
Rev. 04, 03/2015
87
7.2.1
Calibration
In the
"Channel Overview" menu
, click on
Settings | Calibration (
Fig. 54
)
in order
to reach the "Calibration" submenu.
1
Display of the background level
2
Button for changing the background level
3
Button for deleting, adding or changing a calibration point
4
Measurement range input field
5
Entry and display of properties
6
Display of the calibration point - Fill level 50%
7
Display of the calibration point - Fill level 100%
Fig. 55
"Channel Overview" Menu, Settings - Calibration
The calibration submenu serves for the adaptation of the measurement system to
the relevant environmental conditions like e.g. the thickness of the mould, the ac-
tual radiation activity or the adaptation of the background radiation (background
level). The background radiation can vary from location to location.
You can also carry out a multi-point calibration. This is indispensable for operation
with an activity-optimised radiation source (AOS) for factoring in the varying rises
of the calibration profile between 0% - 50% and 50% - 100%, also leading to in-
creased accuracy with commonly designed sources of radiation.
Summary of Contents for castxpert LB 452
Page 2: ......
Page 4: ......
Page 10: ......
Page 14: ...c 2 Safety 14 Rev 04 03 2015 b ...
Page 22: ......
Page 40: ......
Page 120: ......
Page 138: ......
Page 140: ...c 10 Decommissioning 140 Rev 04 03 2015 b ...
Page 146: ...c 11 Technical Information 146 Rev 04 03 2015 b 11 4 c LB 452 Connections Rear Overview ...
Page 151: ...c 12 Declaration of Conformity b Rev 04 03 2015 151 12Declaration of Conformity ...
Page 152: ...c 12 Declaration of Conformity 152 Rev 04 03 2015 b ...
Page 155: ...c 13 Index b Rev 04 03 2015 155 ...