8.2.4
GAMMAcast Control
Code
Text
Description
Correction
Class
501
No detector
found
No compatible GAMMAcast de-
tector found
Connect a GAMMAcast
detector
Ensure the correct ca-
bling is in place
Carry out software up-
date
Contact service
Error
502
Detector com-
munication
Communication error
GAMMAcast - Disruption on
the line
Ensure the correct con-
nection of the screen
Replace cable
Contact service
Warning
503
Detector inter-
nal error
GAMMAcast or ECcast in error
state (Error details and confir-
mation in the GAMMAcast er-
ror menu)
Operating Instructions
GAMMAcast detector or
ECcast system
Error
504
Detector inter-
nal warning
GAMMAcast or ECcast system
in warning status (Error details
and confirmation in the error
menu of GAMMAcast)
See operating instruc-
tions GAMMAcast de-
tector or ECcast system
Warning
505
Detector temp
too high
Detector temperature is/was
too high
Ensure sufficient cool-
ing
Warning
506
Detector
Raw Value
Variation discovered between
the pulses received by the
GAMMAcast and the pulses
transmitted via the communi-
cation
Check cabling
Contact service
Error
Summary of Contents for castxpert LB 452
Page 2: ......
Page 4: ......
Page 10: ......
Page 14: ...c 2 Safety 14 Rev 04 03 2015 b ...
Page 22: ......
Page 40: ......
Page 120: ......
Page 138: ......
Page 140: ...c 10 Decommissioning 140 Rev 04 03 2015 b ...
Page 146: ...c 11 Technical Information 146 Rev 04 03 2015 b 11 4 c LB 452 Connections Rear Overview ...
Page 151: ...c 12 Declaration of Conformity b Rev 04 03 2015 151 12Declaration of Conformity ...
Page 152: ...c 12 Declaration of Conformity 152 Rev 04 03 2015 b ...
Page 155: ...c 13 Index b Rev 04 03 2015 155 ...