6 Maintenance and Repair
Rod source shields LB 8300
40
64217BA2 Rev. 00, 10/2017
To replace the source, proceed as follows:
1.
Prepare the transport shield and open it.
2.
Remove the old source from the shield and insert the old source in the
transport shield, see chapter 6.5.1.
3.
Check the shield for damage and wear and repair it if required
(see chapter 6.2).
4.
Remove the type plate of the old source from the shield and rivet on the
new type plate.
5.
Install the new source in the shield, see chapter 6.5.2.