LED surgical lights, single and combination lights CHROMOPHARE
®
E 778, E 668, E 558
Connecting the electricity
63
11
Connecting the electricity
NOTE!
Electrical installation must comply with the IEC
60364-710 European standard and the current
local standards.
Power switches and fuses must be installed by the
customer (for fuse values
appendix
) or the fuses
and fuse holder delivered must be used.
The cross-section of the connecting lines between
the fuses and the transformers must be at least
1.5 mm².
NOTE!
Detailed electrical plans may be requested from
technical service (
address on the inside of the
back cover of the manual).
NOTE!
Pay attention to the fuse ratings on the fuse
holders!
NOTE!
Fuses must be UL and IEC 60127 approved.