LED surgical lights, single and combination lights CHROMOPHARE
®
E 778, E 668, E 558
Service Menu
95
Fig. 62: Confirming menu item 3
3.
By pressing the
Standby
button, menu item 3
is confirmed.
Thus the address appointed via the DIP-switches on the keypad
PCB will be send to all connected modules.
NOTE!
The light head which is about to be addressed
must not be connected with another light head
via CAN-Bus.
For dual control lights the wall control must be
addressed manually first.
For wall control lights all light heads not about to
be addressed must be disconnected from CAN-
Bus at the swivel arms. The addressing is done via
wall control.