28
4.10
Engineering Unit Menu: Setting up the Engineering Units
Engineering Unit Menu: Allows the user to select various parameters that
relate to the sensor type used, as well as the unit(s) displayed.
Sensor Type, sensor Unit (SU) Sensitivity mV/SU Display unit (DU)
dB reference (DU), Int/Diff, weighting, User defined unit
Detection, HP Filter, Unit detection table
The Engineering Unit menu allows the user to specify the type of sensor you
are connecting to, the units to be measured, the units to be displayed and
weighting that is to be used as well. The Impaq Elite™ is designed to support
a wide range of sensors from Accelerometers, to 200V microphones.
From the pull-down menu, select: Select Sensor Type, Sensor Unit, Sensitivity
mV/SU, Display Unit (DU) dB Ref (DU), Int. / Diff., Weighting, User Def. Unit.
Sensor Type
: Choose Off, Acceleration, Velocity, Displacement, Force or
Sound. If Off is selected, then Volts will be the defined sensor unit.
Sensor Unit (SU):
Choose the unit that is associated with each sensor type.
Sensor Units will only show what is available for the sensor type shown below
in the table.
Acceleration
Gs, Gal, m/s2
Velocity:
Mm/s, in/s, cm/s
Displacement
(mil), (um)
Force
Nt, Grams
Sound
Pa
Off
Volts
Summary of Contents for Impaq Elite