8.2
(2.4.3) Dig. input
Select the digital inputs of the EDS to be configured:
Afterwards, make the settings for the selected digital inputs of the EDS.
8.2
(2.4.3.1) Mode
The operating mode for the digital input can be set to the values below. For a description, refer
to
"Setting the EDS mode (isoxx1685DP only)" on page 33
. The following can be selected
:
8.2
(2.4.3.2) t(on)
The response time t(on) after a switch-on signal can be set between 100 milliseconds and 5 mi-
nutes. For a description, refer to
"Setting the EDS mode (isoxx1685DP only)" on page 33
.
8.2
(2.4.3.3) t(off)
The response time t(off) after a switch-off signal can be set between 100 milliseconds and 5
minutes. For a description, refer to
"Setting the EDS mode (isoxx1685DP only)" on page 33
.
8.2
(2.4.3.4) Function
8.2
(2.4.4) Device settings
8.2
(2.4.4.1) System type
Adjust the EDS to the IT system to be monitored.
•Select all
All digital inputs are selected.
•No selection
No digital input is selected.
•Invert selection
The current selection is inverted.
•Dig. input 1 (BS 2/1)
A single digital input is selected.
•Dig. input 2 (BS 2/2)
•Active high
An event is carried out on the falling edge of the digital input (low to high).
•Active low
An event is carried out on the falling edge of the digital input (high to low).
•off
Digital input without function
•TEST
Device self test
•RESET
Reset of fault and alarm messages
•Select all
All devices are selected.
•No selection
No device is selected.
•Invert selection
The current selection is inverted.
•BMS bus 2 (1-12)
•DC
DC system
•AC
Single-phase AC system
•3AC
3AC system