17
DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
2
Doctor Unit Section
Main Switch
Lit in green
ON OFF
Program version NO.
BELMONT Ver.
00
♪
Belmont 0
Main Switch (Unit Section)
• When the main switch is turned on, the MAIN LED on the
doctor control panel lights up in green.
This state permits motion of the unit, chair and dental light..
The unit can be moved even while the chair is OFF.
• When the main switch is turned on, all the lamps on the
doctor control panel light up, and the following is displayed
on the indicator.
• A few seconds later,
"
BELMONT
" is displayed and all the functions can be operated.
BELMONT
EEPROM READ ERR
If any error arises to the memory, the buzzer sounds and
[EEPROM READ ERR] is displayed on the indicator.
Once turn off the main switch and then turn on the main
switch again. If an error message is still displayed, please
contact your dealer or our office.
* This doctor unit section is actuated by compressed air.
Therefore, ensure that the air compressor is working.
Syringe
Syringe
NO.4
NO.3
NO.2
NO.1
NO.1 NO.2
NO.3 NO.4
When the main switch is turned on, whether handpieces
are accurately fitted in instrument holders or not is checked
automatically. If any handpiece is not accurately fitted in its
instrument holder at this occasion, the subject holder number
(counted as No. 1, 2 and so forth from facing left excluding
syringe) is displayed on the indicator and the buzzer sounds
for a few seconds. Accurately fit the handpiece in the
displayed holder.
If the handpiece is accurately fitted while the buzzer sounds
(for a few seconds), the buzzer stops, and the section become
operative.
If the buzzer stops in the mid-way of the work, turn off the
main switch, accurately fit the handpiece and turn on the
main switch again about 5 seconds later. Then, use of the
section is permitted.