7. Preventive Maintenance/System Operational Check-Out
7-86
B.
ROUTINE MAINTENANCE
1.
Clean and/or Disinfect Exterior
Clean the outside surfaces of the system and inside the door after each use.
a.
Turn the pump to STANDBY and unplug the power cord.
b.
Wipe the surface with a cloth moistened with water or isopropyl alcohol.
Note:
Avoid the use of acetone or other solvents that might damage the
surface.
c.
To remove dried blood and disinfect the pump, clean them with hydrogen
peroxide or a mild bleach solution and dry.
d.
Also clean around the door hinges, making sure the door is pushed all the
way down inside the hinges.
2.
Fluid Out and In-Line Air Detectors
Keep the fluid out and air detectors clean and dry. If they become dirty or wet,
clean with a moistened Q-tip and dry. Air detector surfaces are delicate. Use
care when carrying out this procedure.
3.
Power Cord
Inspect the power cord along its length and connectors for cuts and breaks.
Replace power cord if damaged (replacement part # 118-00096 US and part
#118-00085 International).
4.
Temperature Probes
Keep the IR temperature sensors clean and dry. If they become dirty or wet,
clean with a moistened Q-tip and dry. Use care not to damage the IR sensor
surfaces.
5.
Fan Guards
Inspect the fan guards, on the bottom of the unit, for debris that might impede air
flow. Remove guards by unscrewing the 4 retaining screws and clean, with soap
and water, if necessary. Make certain the guards are not damaged (replacement
part #399-00033). Let the fan guards dry before reinstalling.