8. Alarms and Troubleshooting Guide
8-118
ALARM MESSAGE
POSSIBLE CAUSE
SYSTEM RESPONSE
OPERATOR ACTION
POWER MODULE
OVERTEMP
“POWER MODULE
OVERTEMP FAULT.
KEEP THE AIR INTAKE
AT THE BOTTOM OF
THE MACHINE CLEAR.
PLEASE WAIT FOR
SYSTEM TO CORRECT
THE PROBLEM. SLOW
DOWN THE FLOW RATE
OR USE WARMER FLUID
IF THE PROBLEM
PERSISTS.”
Power driver module
overheating. Temperature
inside the cabinet reaches
70
o
C.
Stop pumping and
heating. Sound alarm
and display message.
Valve in the
recirculate position.
Power goes low and
the fan will go to
maximum speed to
cool the system.
Make certain that the fan air vents at the bottom of the machine are not
blocked.
Wait for unit to correct problem. Display will return to Infuse screen when
the error clears.
Press MUTE to silence the alarm.
Power down and service machine if error persists:
•
Remove door and clean around the door hinges. Push the door back,
making sure it seats all the way down in the hinges.
•
Position system away from heat sources.
PUMP FAULT
“PUMP FAULT, CHECK
PUMP FOR BLOCKAGE.
RESTART SYSTEM OR
PRESS RETRY. SERVICE
MACHINE IF ERROR
PERSISTS.”
Pump failure
Pump speed feedback
encoder failure.
Pump runs out of control
or does not move when
requested.
Stop pumping and
heating. Sound alarm
and display message.
Valve to the
recirculate position.
Check that pump turns freely and head is clean. Pump tubing is seated
properly on the roller pump.
Press Retry to try again.
Power down and service machine if error persists:
•
Remove door and clean around the door hinges. Push the door back,
making sure it seats all the way down in the hinges, or
•
Replace pump motor, or
•
Replace roller pump.
VALVE FAULT
“VALVE FAULT, CHECK
VALVE FOR BLOCKAGE.
RESTART SYSTEM AND
RETRY. SERVICE
MACHINE IF ERROR
PERSISTS.”
Valve failure
Valve position sensor
malfunction. Valve did not
move to the desired
position in 2 seconds.
Stop pumping and
heating. Sound alarm
and display message.
Check that the valve is not blocked.
Restart the system and try again.
Power down and service machine if error persists – replace valve motor, or
Daughter board, or the Computer board.
CAUTION:
In a valve fault condition and fluid is in the disposable set, keep the patient
line clamped closed when opening the door to prevent uncontrolled fluid
flow.