5. Removal and Replacement of Major Components and Modules
5-64
T.
MEMBRANE PANEL
(
Reference Drawing:
403-00107)
Removal
1.
Remove the instrument cover. (See Step A Removal, 1 – 7.)
2.
Remove 1X pan head Phillips screw & 1X flat washer attaching
Membrane Panel Ground Strap to Support Housing.
3.
Remove tape from Membrane Panel Ground Strap.
4.
Remove Membrane Panel Cable from CPU PCB.
5.
Cut silicone from perimeter of Membrane Panel.
6.
Remove Membrane Panel by pressing upward, from the inside of
the Support Housing, along the sides and in the corners.
Installation
1.
Remove protective backing from back surface of Membrane Panel
and adhere to pocket in Support Housing.
Caution
; do not touch
back surface of Membrane Panel.
2.
Reattach Membrane Panel Ground Strap to Support housing with
1X flat washer and 1X pan head Phillips screw.
3.
Apply tape over Membrane Panel Ground Strap to secure in
place.
4.
Reattach Membrane Panel Cable to CPU PCB.
5.
Apply silicone around the perimeter of the Membrane Panel on the
top surface and bottom surface, including around and under cable
and ground strap. Allow silicone to dry for at least 1 hour.
6.
Reinstall the instrument cover. (See Step A Installation, 1 – 5.)