BELIMED WD425E Operating Manual Download Page 20

 

Operating Manual WD 425E

  

 

 

 

 

Rev. 1.

1

, LS                                                                                 Page 20 of 34 

  78563_E rev 1.1.doc 

4.    Program Launch and Process Sequence 

 
!IMPORTANT NOTE! 

• 

Periodic instruction sessions for users is designed to assure cleaning  quality. 

• 

Avoid prolonged decontamination film drying time of medical products. 

• 

Prior to machine type processing, drive elements for drill bits and milling tools or complex 

medical products need to be disassembled. They must be pre-treated by mean of ultra sound or 

manual brushing. Follow the respective manufacturer’s instructions. 

• 

Hollow bodies for drive elements or MIC instruments generally need connection of hoses at the 

wash good racks. This in order to assure decontamination and disinfection of inner surfaces of 

the respective medical products. 

 

 

1. 

Check liquid level of detergent. If needed, replenish container.  

2. 

Open supply shut off valve and energize machine by main switch. 

3. 

Push 0-I switch. Display shows „Program ready “ 

4. 

Press door key to open wash chamber door, remove insert rack. 

5. 

Load appropriate rack only with wash goods. Be sure that water is always capable of running 
of wash goods and that all wash arms cane freely rotate

6. 

Drain all containers

, remove plugs, corks, tags, sealing wax remnants etc. Wash good utensils may 

not be piled inside each other and thus cover each other. 

7. 

The wash arms must not be blocked by protruding (top or bottom) wash good elements. Carry 

out manual rotation check.

 

8. 

Wash goods must not stick out of the racks. When pushing rack into machine, protruding sharp ob-
jects might otherwise be harmed. 

9. 

When pushing rack into machine, it is mandatory that water docking elements match those of 

the machine!  

10. !DANGER! 

 

A fully loaded rack may weigh up to 100 kg. Caution, a falling rack may put operators at risk of 

injury. 

11. 

With active batch ID, the barcode reader may identify the user, the wash goods rack and also the 
batch content (endo basket, sieve trays). 

12. 

Fully insert wash goods rack until docking elements for supply are felt snapping into their counter-
parts of the machine. 

13. 

With the WD 425, a clicking sounds must be audible and display must show „Program ready “. 

14. 

The machine is now ready for launch. 

15. 

Press the desired program key. Check if water is received by the wash chamber. Display initially 
shows the program name, subsequently the respective program status. 

16. 

During the program run, the display will keep you informed about the current program step, the cur-
rent temperature inside the wash chamber and about the remaining program run time. 

17. 

During the entire wash cycle, the wash chamber door remains locked. 

18. 

If during the program cycle the program is stopped by means of the I/O button, the door remains 
locked nevertheless . Instead of program, the display shows „Door locked „. The door can be unlo-
cked only at the end of a full and successful program cycle. The door also remains locked as long as 
the level sensor shows >0.5V (water in machine). 

Summary of Contents for WD425E

Page 1: ...User Operating Manual WD425E Copyright F Gehrig AG Ballwil Switzerland 2005 Art Nr 78563_E...

Page 2: ...ndoscopes 11 1 3 3 Leak Test for Endoscopes 12 1 3 4 Manual Leak Test 12 1 3 5 Automatic Leak Test 13 1 3 6 After Endoscope Processing 13 1 4 Instrument Rack 14 1 5 AN Rack 14 2 Operation Keyboard 15...

Page 3: ...external 24 6 6 The Machine does not wash properly 24 7 Machine does not run 24 8 Error Indication at Time of Process Interrupt 25 9 Activate Modem Connection 32 10 Wash Chamber Cleaning and Maintena...

Page 4: ...result in damage to the machine The operation must be handled by prop erly instructed and trained personnel CAUTION Cleaning and disinfection performance A cleaning assignment is assured only if a va...

Page 5: ...s must not be ingested prevent eye and skin contact Check if the appropriate additive is being replenished or exchanged Dosing must be verified following each container exchange Detergents ad their vi...

Page 6: ...provide pull relief Forceful pulling on components may lead to electric shock Prior to installation of addi tional equipment or replacement of components machine must be rendered powerless Caution Onl...

Page 7: ...od and proteins foam formation in the CDM Impediment of decontamination is given by Pretreatment with aldehyde disinfection media Pretreatment with alcoholic solutions Dropping antiseptic solutions on...

Page 8: ...hemical analyses of the water must be expanded by micro biological investigations to keep opportunistic and ill making germs in check 1 2 3 Appropriate Rinse oriented Loading Influence of loading prop...

Page 9: ...st Attach water protective cap on video scopes to guard electrical contacts Deposit endoscope in a bowl containing cleaning solution Remove all valves and distal caps place them in cleaning solution C...

Page 10: ...y endoscope is ready for renewed application with patients Storage of endoscope must be done in totally dry dust protected preferably in suspended position in an endo scope cabinet Store valves dust f...

Page 11: ...osition Carefully close pressure chamber cassette When closing lid be sure to prevent any damage to the supply end Place supply end on wire spiral helix Connect adapter with supply components for leak...

Page 12: ...ed without water contact Build up of test pressure Setting phase 10 seconds Air disperses evenly within the endoscope Test phase Holding inside pressure in the endoscope within a predetermined range a...

Page 13: ...r minute i e 52 mV error 171 or 176 is shown As the pressure drops the dots disappear according to pressure values Venting applies also after endoscope protection The endoscope is being vented In case...

Page 14: ...n connection with special proce dures and retainer equipment Simple tools and accessories may be proc essed just like surgical instruments 1 5 AN Rack Inappropriate loading improperly placed anes thes...

Page 15: ...is like program recipe and setup calibration 2 Beeper ON OFF at program end and in case of error an acoustic signal is sounded 3 Printer ON OFF 4 Shift key programs P7 12 Program keys PT 1 2 3 4 5 6...

Page 16: ...uption value is ignored Cleaning Current step 36 C Temperature of wash liquid or of drying air at input wash channel 3 3 User Identification Importing Names N by way of the keyboard or importing Name...

Page 17: ...keys from 11 66 or a text by barcode scanner The respective number e g 36 or name is passed on to the printer or to the digital documentation system 8565 and assigned to the respective batch A progra...

Page 18: ...t No ID will be passed on The reading process may be terminated at any time by the IO key All inputted data will be erased At he end always conclude with N 0 On the display indication of number of bat...

Page 19: ...rface RS 485 X32 to show on display As soon as the first number is input the right hand endoscope is interrogated Display shows Read Content No Read batch content N R If no endoscope is placed in spot...

Page 20: ...ottom wash good elements Carry out manual rotation check 8 Wash goods must not stick out of the racks When pushing rack into machine protruding sharp ob jects might otherwise be harmed 9 When pushing...

Page 21: ...not be unlocked A program must always be successfully completed Suppressing wash liquid drainage according to 18 ISSG Infection Protection Law BGA RKI Drainage of wash liquid in case of program inter...

Page 22: ...ion by qualified service personnel This procedure must be repeated periodically All validating provisions must be heeded Periodic hygiene tests are to guarantee that the wash goods have been cleaned a...

Page 23: ...y display shows Door lock This is a precaution for the user In case of malfunction e g if wash pump cannot be shut off door must remain locked Remedy Upon successful completion of program run the door...

Page 24: ...connections open 2 Is there sufficient cleaning and disinfecting additives in the storage containers 3 Are dosing lines open and permeable 4 Have wash goods been properly loaded on appropriate racks 5...

Page 25: ...tive program for a period of minimum 5 sec onds control system cuts the running program off and displays an error signal Interrogation input S211 only with media M1 4 M7 9 90 seconds after filling pha...

Page 26: ...running program under error indication Monitoring of limit value input PR during program cycle Endo protection Error code 123 130 Temperature sensor in wash chamber is defective or is not connected I...

Page 27: ...e Clean or replace Level measuring system P1 is not tight Check hoses and connections S 110 door switch does not close Check door switch and its position No cold water CW Error code 141 142 Working le...

Page 28: ...el change during drain cycle of wash cham ber If during drainage by means of drain pump K31 the liquid level does not change during a predetermined period td 3 Min down to level 0 7 V empty level 0 2V...

Page 29: ...program run in a step phase a dosing procedure with flow meter 1 impulses is programmed at least 10 impulses possibly only air is pumped at first must be counted during dosing with dosing pump M 21 w...

Page 30: ...out of order Replace Cable connections are burned through Replace connector Drain valve leaks Clean replace membrane Suction system has excessive suction Reduce negative pressure Exhaust damper remai...

Page 31: ...2 74 2 34 V has not been reached within 10 minutes Cause On right hand endoscope at endoscope adapter or on system within the machine a signifi cant leak has been detected Compressor or test valves of...

Page 32: ...eriodic service can deliver immaculate results Daily checks wash chamber visual check clean sieve system including course sieve must be cleaned remove foreign objects reposition all sieves with great...

Page 33: ...assignments validation and counseling is at your disposal IMPORTANT ADVICE The housing must not be hosed down The machine is NOT protected against water jets WARNING Hand protection is required to cle...

Page 34: ...x 41 71 644 86 00 Email sag sauterag com For maintenance and Service in all other countries BELIMED INFECTION CONTROL Dorfstrasse 4 CH 6275 Ballwil Switzerand Telephone 41 41 449 78 88 Telefax 41 41 4...

Reviews: