Operating Manual WD 425E
Rev. 1.
1
, LS Page 19 of 34
78563_E rev 1.1.doc
3.6 Endoscope Identification
With active endoscope detection, with door open, display shows “Interrogation Batch Content” instead of
„Door open “. With door button, leak test may be started simultaneously:
Read Content N° / Input batch content
Endo L = - - ?
With barcode reader or keyboard, read batch content left hand endoscope or keyboard.
SW reads a 2-digit number (12 CHAR) with 12 characters , e.g. 000000000012.
Read Content N° / Read batch content
Endo
L
=
12
The number read (e.g. 12) or the name is passed on to the printer RS232 X 31 or to the interface RS 485 X32 to
show on display.
As soon as the first number is input, the right hand endoscope is interrogated. Display shows:
Read Content No. / Read batch content
N° R = - - ?
If no endoscope is placed in spot, by N° 66, the next interrogation will be pursued or reading of right hand endo-
scope may be terminated.
Read Content No. / Read batch content
N° R = No identification
Instead of N° 66, printer shows “No ID”
Following reading in the second endoscope, display shows
Program ready
Door open
If there is no input and a program is started, display shows „No ID “.
If unintentionally an incorrect number is entered, pushing the I/O button will disrupt the entire procedure. All previ-
ously entered data will thus be lost.
Program cut off
13.02.2003 11.15
With active endoscope ID, printer will show the line „Batch content.