Operating Manual WD 425E
Rev. 1.
1
, LS Page 11 of 34
78563_E rev 1.1.doc
1.3.2 Placing of water tight, flexible Endoscopes
First, remove valves, distal caps etc.
Open pressure chamber cassette.
Insert distal end of endoscope in spiral shaped Teflon hose until the hand part can be placed in wash box.
Position hand portion of endoscope such as to have supply end fit in the respective notch of the wash box, then
lock in position.
Carefully close pressure chamber cassette. When closing lid, be sure to prevent any damage to the supply end.
Place supply end on wire spiral helix.
Connect adapter with supply components for leak test. For this, on endoscope side, the connection used for the
manual leak test is to be used. The other end of the adapter hose is connected on machine side in the wash
chamber at the ceiling, right and left side respectively.
The supply end is in-
serted in the cutout of
the endo box, held in
place and cover care-
fully closed.
The
pressure
test
adapter is connected with
the appropriate location
in the wash chamber.