BELFORT
INSTRUMENT
Proprietary Information
Runways:
_____________.
Sensor Elevation:
_____________.
Field Elevation:
_____________.
Magnetic Variation:
_____________.
FBO Name:
_____________.
FBO/Unicom Frequency:
_____________.
Site Password:
_____________.
RF Transmit Power:
_____________.
Visibility Sensor Type:
_____________.
Visibility Sensor Baud Rate:
_____________.
Configuration Values saved in EEprom:
Pass______ Fail:_____.
Complete Air Nav Document attached to this sheet:
Yes ______ No:_____.
DCP Enclosure cover screws fully tightened:
Yes ______
No:_____.
3.
Barometric Pressure Sensor (refer to Maintenance Manual section 5.2)
DCP S/N:
Reference
Belfort AWOS
Acceptable
Adjustment Value
_________
Standard (A) Reading (B) Tolerance
Required?
After
PN: _______
Adjustment
BP1 _________
_________
±0.02 in Hg
Y ___ N ___ _________
BP2 _________
_________
±0.02 in Hg Y ___ N ___ _________
Review the past year maintenance log to see if the DCP (with pressure sensors) had been
replaced. If it was replaced, indicate that fact here.
DCP Replaced Yes
No
From the maintenance logs, review the previous measured pressure sensors values and
document any change in the measured sensor values since the last altimeter check.
Sensor Drift
BP1_________ BP2 _________ < +- .02 in Hg.
Record any configuration difference(s) identified _____________________________
4. Visibility Sensor (refer to Maintenance Manual section 5.4)
Visibility Sensor PN: __________________
Serial Number: __________
A. Clean visibility and ambient sensor windows
Completed
B. Perform sensor calibration if the ExtCo difference is > +-5%:
Acceptable
Sensor
Reference
% Diff
Tolerance: ± 5%
ExtCo
ExtCo
Initial
__________ __________ ______
Pass
Fail
After Cal.
__________ __________ ______
Pass
Fail
5. Tower/DCP Mounting Assembly Visual Inspection:
A. Check for obvious mechanical damage to system and repair or replace as needed
Summary of Contents for AWOS
Page 17: ... 14 BELFORT INSTRUMENT Proprietary Information Figure 2a Securing the Tower for Maintenance ...
Page 49: ... 46 BELFORT INSTRUMENT Proprietary Information Figure 9 Temperature Humidity Sensor Removal ...
Page 51: ... 48 BELFORT INSTRUMENT Proprietary Information Figure 10 Visibility Sensor Removal ...
Page 60: ...BELFORT INSTRUMENT Proprietary Information Appendix A Belfort AWOS Maintenance Forms ...
Page 66: ...BELFORT INSTRUMENT Proprietary Information This page intentionally left blank ...
Page 74: ...BELFORT INSTRUMENT Proprietary Information Appendix C UHF Transceiver Specifications ...
Page 88: ...BELFORT INSTRUMENT Proprietary Information Appendix F DCP Site Configuration Procedure ...