BELFORT
INSTRUMENT
Proprietary Information
5.
Barometric Pressure Sensor (refer to Maintenance Manual section 5.2)
DCP S/N: Reference
Belfort AWOS
Acceptable
Adjustment
Value
_________ Standard(A)
Reading (B)
Tolerance
Required?
After Adjustment
BP1
_________
_________
±0.02 in Hg
Y ___ N ___
_________
BP2
_________
_________
±0.02 in Hg
Y ___ N ___
_________
Review the past year maintenance log to see if the DCP (with pressure sensors) had been replaced. If
it was replaced, indicate that fact here.
DCP Replaced
Yes
No
From the maintenance logs, review the previous measured pressure sensors values and document any
change in the measured sensor values since the last altimeter check.
Sensor Drift
BP1_________ BP2 _________ < +- .02 in Hg.
Record any configuration difference(s) identified _____________________________
6. Visibility Sensor (refer to Maintenance Manual section 5.4)
Review the past years maintenance log and verify that no replacement of the Visibility sensor was
made. If a replacement was made, verify that P/N and Revision installed is acceptable to current
authorized VDD baseline.
Yes
No
Record any configuration difference(s) identified _____________________________
7. Perform the Annual Wind Speed Sensor confidence validation (refer to Maintenance Manual section
5.5)
Pass
Fail
Review the past years maintenance log and verify that no replacement of the Wind Speed sensor was
made. If a replacement was made, verify that P/N and Revision installed is acceptable to current
authorized VDD baseline.
Yes
No
Record any configuration difference identified __________________________________
8. Perform the Annual Wind Direction Sensor confidence validation (refer to Maintenance Manual
section 5.6)
Pass
Fail
Review the past years maintenance log and verify that no replacement of the Wind Direction sensor
was made. If a replacement was made, verify that P/N and Revision installed is acceptable to current
authorized VDD baseline.
Yes
No
Record any configuration difference identified ______________________________
Summary of Contents for AWOS
Page 17: ... 14 BELFORT INSTRUMENT Proprietary Information Figure 2a Securing the Tower for Maintenance ...
Page 49: ... 46 BELFORT INSTRUMENT Proprietary Information Figure 9 Temperature Humidity Sensor Removal ...
Page 51: ... 48 BELFORT INSTRUMENT Proprietary Information Figure 10 Visibility Sensor Removal ...
Page 60: ...BELFORT INSTRUMENT Proprietary Information Appendix A Belfort AWOS Maintenance Forms ...
Page 66: ...BELFORT INSTRUMENT Proprietary Information This page intentionally left blank ...
Page 74: ...BELFORT INSTRUMENT Proprietary Information Appendix C UHF Transceiver Specifications ...
Page 88: ...BELFORT INSTRUMENT Proprietary Information Appendix F DCP Site Configuration Procedure ...