28
With the
“Manual” start mode
Perform the zeroing if necessary by pressing the
key
.
Load the sample tray on the weighing star-
shaped item and switch off the heater. Then
press the key
to confirm.
After resetting the sample tray, load the
substance to be analyzed on the tray.
With the
“Automatic” start mode
Perform the zeroing if necessary by pressing the
key
.
Load the substance to be analyzed on the
sample tray and switch off the heater.
With the
“Manual” start mode
Perform the zeroing if necessary by pressing the
key
.
Load the substance to be analyzed on the
sample tray and switch off the heater. Then
press the key
to confirm.
N.B. The minimum weight of substance
allowed for the drying cycle varies according
to the scale resolution:
Resolution 0.01g => Min 500mg
Resolution 0.001g => Min 500mg
Resolution 0.0001g => Min 50mg
After closing the heater, the analysis cycle will be
started.