Operation METPOINT® BDL compact
METPOINT® BDL compact
59
Fifth step:
define the register
Main menu
Settings
Sensor settings
A1
Va
Use
Sixth step:
enter the Modbus parameters
The following parameters can be retrieved via the corresponding registers:
Register
Designation
Address of
register
Reg. format
Unit
Scal.
Va
Temperature
1216
[HR ]R4
°C
No scal.
Vb
Rel. humidity
1152
[HR ]R4
% rH
No scal.
Vc
Dew/freezing
point
1536
[HR] R4
°C t
d
No scal.
Vd
Dew point
1472
[HR ]R4
°C t
d
No scal.
The definition of other registers is imple-
mented in the same manner.
The settings of the reg./data format are the
same for all the registers.
The entry of the Modbus parameters is imple-
mented via the white buttons (1) – (4).