background image

13

6.4

Position of the decimal point:  

dP

A dummy decimal point can be positioned between
any of the digits or it may be absent. To position
the decimal point select

dP

from the menu and

press

(

.

The decimal point can be moved by

pressing the

&

or

*

push button. If a decimal

point is not required it should be positioned beyond
the most or least significant digit.

When

positioned as required press the

)

button to enter

the selection and return to the configuration menu.

6.5

Calibration using an external
current source:  

CAL

This function enables the zero and span of the
indicator to be adjusted using an external
calibrated current source.

When used with an

accurate traceable current source this is the
preferred method of calibration.

Zero is the indicator display with 4mA input
Span is the indicator display with 20mA input

To calibrate the indicator select

CAL

from the

configuration menu and press

(

. The indicator

will display

2Ero

which is a request for a 4mA input

current. Adjust the external current calibrator to
4.000mA and again press

(

which will reveal the

current zero display.

The flashing digit of the

indicator display can be changed by pressing the

&

or

*

buttons, when set as required pressing

(

will transfer control to the next digit. When all

the digits have been adjusted, press

)

to enter

the new zero and return to the 

2Ero

 prompt .

Pressing the

*

button will cause the indicator to

display

5PAn

which is a request for a 20mA input

current. Adjust the external current calibrator to
20.000mA and again press

(

which will reveal

the existing span display. The flashing digit of the
indicator display can be changed by pressing the

&

or

*

buttons, when set as required pressing

(

will transfer control to the next digit. When all

the digits have been adjusted press

)

to enter

the new span and return to the

5PAn

prompt.

Finally

press

)

again to return to the

configuration menu.

Notes:

a. The indicator input current must be adjusted to

the required value before the zero and span
functions

are

entered

by

pressing

the

(

 button.

b. Indicators may be calibrated at currents other

than 4 and 20mA, within the range 3.8 to
21.0mA providing the difference between the
two currents is greater than 4mA.

If these

conditions are not complied with, the indicator
displays 

FaiL

 and aborts the calibration.

c. If the zero current is greater than the span

current the instrument will be reverse acting i.e.
an increasing input current will cause the
display to decrease.

6.6

Calibration using internal reference: 

5Et

Using the

5Et

function the indicator can be

calibrated without the need to know the value of
the 4/20mA input current, or to disconnect the
indicator from the 4/20mA loop.

When using the

5et

function the indicator’s internal

reference is used to simulate a 4mA and 20mA
input current.

Zero is the display with a simulated 4mA input
Span is the display with a simulated 20mA input

To calibrate the indicator display select

5Et

from

the configuration menu and press

(

.

The

indicator will display

2Ero

, pressing

(

again will

reveal the current display at 4mA.

The flashing

digit can be adjusted by pressing the

&

or

*

buttons, when the flashing digit is correct pressing

(

will transfer control to the next digit. When all

the digits have been adjusted, press

)

to return

to the 

2Ero

 prompt.

To adjust the display at 20mA, press the

*

button

which will cause the indicator to display

5PAn

,

pressing

(

will then reveal the indicator’s existing

display at 20mA.

The flashing digit can be

adjusted by pressing the

&

or

*

buttons, when

the flashing digit is correct pressing

(

will transfer

control to the next digit. When all the digits have
been adjusted press

)

to return to the

5PAn

prompt followed by

)

to return to the

5Et

prompt

in the configuration menu.

6.7

Bargraph format and calibration:  

bAr

Only the BA324G and BA324G-SS have a
bargraph.

In addition to a five digit numerical display the
BA324G has a 31 segment analogue bargraph
which may be configured to start and finish
anywhere within the indicator’s numerical display
range.

To configure the bargraph select

bAr

from the

configuration menu and press

(

. The indicator

will display

tYPE

, pressing

(

again will reveal the

existing bargraph justification which can be
changed to one of the following four or five options
using the 

&

 or 

*

 button:

Bargraph justification starts from

LEFt

Left end of display

CEntr

Centre of display

riGHT

Right end of display

Alr5P

Only

 

with

 

alarms 

see

 

section 9.3.14

oFF

Bargraph disabled

When set as required press

)

to return to the

tYPE

 sub-function prompt.

Summary of Contents for BA304G

Page 1: ...Issue 5 19th September 2018 BA304G BA304G SS BA324G BA324G SS intrinsically safe loop powered field mounting indicators Issue 5...

Page 2: ...to a new indicator 7 3 Under and over range 7 4 Lineariser default configuration 8 Maintenance 8 1 Fault finding during commissioning 8 2 Fault finding after commissioning 8 3 Servicing 8 4 Routine ma...

Page 3: ...sting and Certification Ltd and comply with the European ATEX Directive 2014 34 EU The EC Type Examination certificate specifies that under fault conditions the output voltage current and power at the...

Page 4: ...d Body Intertek Testing and Certification Ltd have issued all the models with a common EC Type Examination Certificate number ITS11ATEX27253X This confirms compliance with harmonised European standard...

Page 5: ...face of the instrument assembly It shows the ATEX and IECEx certification information and BEKA associates name and location Non European certification information may also be shown The instrument seri...

Page 6: ...sufficient to drive the indicator plus the interface When a high integrity earth connection is already available a Zener barrier is usually the least expensive option If an earth connection is not ava...

Page 7: ...ing cable screens If field wiring is to be terminated prior to the installation of the indicator assembly the GRP models BA304G and BA324G can be supplied with an optional back box terminal assembly w...

Page 8: ...for joining cable screens 5 6 Units of measurement and tag marking on scale card The indicator s units of measurement and tag information are shown on a scale card which slides into the indicator New...

Page 9: ...e card carefully insert it into the slot on the right hand side of the input terminals as shown in Fig 6b Force should be applied evenly to both sides of the scale card to prevent it twisting The card...

Page 10: ...A324G BA304G SS BA324G SS Access code CodE 0000 0000 Function FunC Linear Linear Display at 4mA 2ero 0 0 0 00 Display at 20mA 5PAn 100 0 100 00 Resolution rE5n 1 digit 1 digit Bargraph start BarLo 0 0...

Page 11: ...11...

Page 12: ...Lin 16 segment adjustable lineariser To reveal the indicator s existing function select FunC from the configuration menu and press If the function is set as required press to return to the menu or pre...

Page 13: ...e two currents is greater than 4mA If these conditions are not complied with the indicator displays FaiL and aborts the calibration c If the zero current is greater than the span current the instrumen...

Page 14: ...on and oFF When set as required press to return to the tARE prompt in the configuration menu 6 10 Security code CodE Access to the instrument configuration menu may be protected by a four digit securi...

Page 15: ...ws direct calibration of the lineariser with an external current source and is the preferred method when traceability is required If the exact system non linearity is unknown this method also allows d...

Page 16: ...16...

Page 17: ...g break point 7 1 1 Example Adding break points to a new indicator When adding break points to a new indicator or to a lineariser following resetting to the factory defaults using the LtAb function de...

Page 18: ...pressing the button The corresponding indicator display at each of the break points can now be defined using the di5P sub function Using the and buttons select the di5P sub function and press which wi...

Page 19: ...ertificate is available If an indicator fails after it has been functioning correctly follow the procedure shown in section 8 1 If this does not reveal the cause of the fault it is recommended that th...

Page 20: ...ered the backlight produces green background illumination enabling the display to be read at night or in poor lighting conditions No additional power supply intrinsic safety interface or field wiring...

Page 21: ...rgency shut down system Both models can be supplied with factory fitted dual solid state single pole alarm outputs Each alarm output may be independently configured as a high or low alarm with a norma...

Page 22: ...simple apparatus This simplifies system documentation and allows the alarm output terminals 8 9 and 10 11 to be connected to almost any intrinsically safe circuit protected by a Zener barrier or galv...

Page 23: ...23...

Page 24: ...tpoint The alarm is activated when the indicator display equals the setpoint See section 9 3 5 Hi Lo Alarm function Defines the alarm function as High or Low See section 9 3 6 no nC Normally open or n...

Page 25: ...for fail safe operation both alarm outputs should be conditioned to be open in the alarm condition nC To check or change the alarm output status select no nC from the alarm configuration menu and pre...

Page 26: ...t having access to the indicator configuration menu Protection against accidental adjustment of the setpoints when the indicator is in the display mode is provided by a separate security code This dir...

Page 27: ...tpoint plus the displayed value to be represented or a low and a high setpoint plus the displayed value to be represented by the bargraph as shown in Fig 18 Fig 18 Displayed value and setpoints on bar...

Page 28: ...d into a panel aperture The kit provides an IP66 seal between the indicator and the instrument panel it is therefore suitable for mounting a stainless steel indicator in a sealed panel enclosure 9 6 B...

Page 29: ...nly persist for a short period Be used with dust in subdivisions IIIA combustible flyings IIIB non conductive dust IIIC conductive dust Having a Minimum Ignition Temperature of Dust cloud 120 C Dust l...

Page 30: ...cifies the following certification codes Ex ia IIC T5 Ga Ex ia IIIC T80 C Da IP66 Ta 40 C to 70 C The specified IECEx gas and dust intrinsic safety parameters are identical to the ATEX safety paramete...

Page 31: ...70 C CL II Div 1 Groups E F G 40 C Ta 60 C CL III Div 1 40 C Ta 60 C USA Zone classification CL I Zone 0 AEx ia IIC T5 Ga 40 C Ta 70 C Zone 20 AEx ia IIIC T80 C Da 40 C Ta 60 C Canada Zone classificat...

Page 32: ...32...

Page 33: ...33...

Page 34: ...34...

Page 35: ...35...

Page 36: ...36...

Page 37: ...37...

Page 38: ...38...

Page 39: ...39...

Reviews: