Recommended distance between portable and mobile communications equipment and
the SENTA PUR L. (Table 206)
The SENTA PUR L is intended for use in an electromagnetic environment with controlled HF
interferences. The customer or the user of the SENTA PUR L can avoid electromagnetic
interference by respecting and observing the minimum distance between portable and mobile
HF telecommunication devices (transmitters) and the SENTA PUR L depending on the rated
output of the communication device as given below.
Rated output
of the transmitter
W
Separation distance depending on the transmitting frequency
in m
150 kHz to 800 MHz
d=0.35√P
80 MHz to 800 MHz
d=1.2√P
800 MHz to
2.5 GHz
d=2.3√P
0.01
0.1
1
10
100
0.04
0.11
0.35
1.1
3.5
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance
d
in meters (m) can be determined using the equation applicable to the
frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in
watts (W) according to the specifications given by the transmitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2:
These guidelines could not apply to all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Summary of Contents for SENTA PUR L
Page 54: ...Appendix Operating manual SENTA PUR L V2 5 E Page 54 of 64 14 2 Dimensions of the SENTA PUR L...
Page 55: ...Appendix Operating manual SENTA PUR L V2 5 E Page 55 of 64 14 3 Declaration of conformity...
Page 63: ...Appendix Operating manual SENTA PUR L V2 5 E Page 63 of 64 Your notes...
Page 64: ...BEKA Hospitec GmbH Am R benmorgen 3 35582 Wetzlar Germany Phone Fax 49 0 6 41 9 22 20 0...