8.11.1
Safety instructions digital scale SL
All persons involved in the setting and the maintenance of the SL scale, must:
•
have been instructed and trained accordingly,
•
have read and understood the operating manual and observe the instructions
contained therein,
•
ensure that the safety instructions are respected and observed.
8.11.2
Cleaning the scale
Wipe the scale off with a damp cloth.
Never use abrasive or corrosive cleaners.
8.11.3
Using the scale
In order to ensure a proper operation of the scale, please observe the following:
•
Prior to Weighing:
Push the on/off button to switch the scale on.
Wait until the scale is automatically tarred to 0.0 kg.
•
Preparing the weighing procedure:
➢
Position the patient/resident onto the SENTA PUR L combination bathing lift.
➢
Make sure that the patient/resident is sitting or lying completely on the
combination bathing lift and that the weighing procedure is not influenced by the
patient/resident partly resting on e.g. a chair, a wheelchair, a bed.
➢
The correct weight of the patient/resident is now shown (in 0.1 kg steps).
Caution:
When calculating the weight, please observe that the proper weight of the clothes of
the patient/resident must be deduced from the displayed weight to determine the
actual body weight.
•
The scale can now be switched off by pressing the on/off button.
8.11.4
Zeroing and Taring
Push the →0/T← button to zero the digital scale prior to a weighing procedure and
without initial load. With this function, you can tare initial loads (sling, stretcher lift
wings, straps) as well. These initial loads must be correctly applied prior to the
weighing and tarred by pushing the button.
Summary of Contents for SENTA PUR L
Page 54: ...Appendix Operating manual SENTA PUR L V2 5 E Page 54 of 64 14 2 Dimensions of the SENTA PUR L...
Page 55: ...Appendix Operating manual SENTA PUR L V2 5 E Page 55 of 64 14 3 Declaration of conformity...
Page 63: ...Appendix Operating manual SENTA PUR L V2 5 E Page 63 of 64 Your notes...
Page 64: ...BEKA Hospitec GmbH Am R benmorgen 3 35582 Wetzlar Germany Phone Fax 49 0 6 41 9 22 20 0...