V1.0K3 PULSE OXIMETER INSTRUCTION MANUAL
11
Do not use batteries of different types together.
Do not use new and used batteries together.
Dispose of batteries in accordance with the local ordinances and regulations.
Regular maintenance of rechargeable batteries is required, such as periodic charge
and discharge.
The charging time shouldn’t be too long, otherwise, it may reduce batteries
performance and shorten batteries’ life.
2.3.2 Charging Batteries through adapter (optional)
To charge batteries, please ensure the installed batteries are Ni-MH ones. Firstly connect
the adapter with the oximeter and secondly with the wall outlet, and then press and hold
the power switch for 3 seconds to power the device on.
A prompt window will appear inquiring you “Are the batteries Ni-MH ones?”. Select “Yes”
to charge them by wall outlet or “No” to abandon charge.
Notes:
1.
To avoid the device from being damaged due to shot circuit, please make sure the
adapter is firstly connected with the oximeter and secondly with the wall outlet. While
disconnected the connection, please make sure the adapter is firstly disconnected from
the wall outlet and then disconnected from the oximeter.
2.
During battery charge, the “Power Auto” and “Brightness” items can not be accessible.
3.
It is not recommended to take a measurement during the oximeter is powered by wall
outlet or being charged, for damages and injuries may be caused to the device or users.
4.
DO NOT charge when the non-rechargeable batteries are installed, otherwise damages
or injuries may be caused to the device or users.
5.
You can cut off the power of the device through unplug the adapter.
6.
It is not recommended to charge batteries while the battery power is sufficient. If you
perform the charging action at this time, the device will detect the battery voltage and
terminate the charging automatically to protect the batteries. In this case the prompt
window will not appear.
3 Install SpO
2
Probe
Set the fixing clip for SpO
2
probe, and then insert the SpO
2
probe to the SpO
2
socket as