Please consider your environment when you dispose of your
one-way batteries or accumulators. They belong in a rubbish
dump for hazardous waste. In most cases, the batteries can
be returned to their point of sale.
Please, comply with the respective valid regulation regard-
ing the return, recycling and disposal of used batteries and
accumulators.
6.0
Technical Data
Display: ..............................................3 digit, digital LCD
Voltage range/ ....................................1 ... 700 V / AC/DC
Resolution: ................................................................1 V
DC/AC voltage detection:..................................automatic
Meas. range select: ..........................................automatic
Polarity indication: ..........................................total range
Accuracy (AC/DC): ..........................± (3% rdg + 3 digits)
Frequency range: ..........................................0 ... 400 Hz
Test current: ......................................................< 3,5 mA
Internal basic load: ......................approx. 2.1 W at 700 V
Peak current: ..................................................Is < 0,25 A
Automatic load: ................for tripping a 10 mA or 30 mA
........................................................RCD circuit-breaker
Auto-Power-On: ........................from approx. 1 V AC/DC
Single-pole phase test: ........................100 V ... 700 V AC
Contactless Voltage Test: ............................50...700V AC
........................Detection in min. 5 cm distance (NYM 3x1,5 7 230V / 50Hz)
Frequency range: ........................................50 ... 400 Hz
Display: ..............................................................acoustic
Two-pole rotay field indication:................100...700 V AC
Frequency range: ............................................50...60 Hz
Display:........................................optical via LCD symbol
Continuity test / diode test: ............................0... 200 k
Ω
Display: ............................................optical and acoustic
Test current: ................................................approx. 5 µA
Resist. meas. range: ........................................1...1999
Ω
Resolution: ................................................................1
Ω
Accuracy: ........................................± (3% rdg + 3 digits)
Execution: ......................................DIN VDE 0680, Part 5
Temperature range:......................................-10°...+55°C
Hight above MSL: ......................................up to 2000 m
Humidity: ..............................max. 85% relative humidity
Protection: ..............................................................IP 50
34