GT-600/GT-800 Instruction Manual
61
Reset of the GT-600/GT-800 Tester
There are many adjustable parameters in the GT-600/GT-800 Tester. If an
operator due to any reason wishes to reset all adjustable parameters to factory-
set values, this operation can be done by performing RESET function.
Switch off the GT-600/GT-800 Tester by using the mains switch (4).
Press and keep pressing the function key SAVE (F5) while switching on the GT-
600/GT-800 Tester by using the mains switch (4).
“RESET” information will be displayed for a while informing the operator the
RESET function has been carried out.
Release the function key SAVE (F5).
The following parameters have been reset.
Function Parameters
R
PE
-
Test current 5 A
-
Limit value 0.3
-
MAN mode
-
No test lead compensation
R
ISO
-
Test voltage 500 V
-
Protection class I
-
Limit value 1.00 M
(PC I)
-
Limit value 0.30 M
(PC I-HEATER)
-
Limit value 2.00 M
(PC II)
-
Limit value 0.25 M
(PC III)
I
SUB
-
Protection class I
-
Limit value 3.50 mA (PC I)
-
Limit value 0.50 mA (PC II)
I
PE(
)
-
Positive mains polarity
-
Limit value 3.50 mA
I
T
-
Positive mains polarity
-
Limit value 0.50 mA
CLAMP (GT-800 only) -
Range 0 ... 60 A
PRCD/CORD (GT-800 only) -
CORD
CORD -
No cable compensation
-
Limit value 0.30
PRCD (GT-800 only) -
No cable compensation
-
Limit value 0.30
-
I
N
= 30 mA
-
MAN mode
PELV/SELV -
PELV
PELV
-
Limit value 50 V
SELV
-
Limit value 50 V
AUTO (GT-800 only) -
Selected AUTO test No. 145 (German region)
MENU -
Operator XXXX
-
Measurement times
-
MAN, MAINS ... 10 seconds
-
MAN, MAINS FREE ... 3 seconds
-
AUTO, MAINS ... 10 seconds
-
AUTO, MAINS FREE ... 3 seconds
-
Region GERMANY
-
Acoustic signal ON
-
Barcode reader mode STANDARD
-
Language GERMAN
-
Contrast 50 %