GT-600/GT-800 Instruction Manual
68
Substitute Leakage Current TRMS (I
SUB
)
Available protection class
I and II
Measurement terminals class I
L/N against (PE on test socket in parallel with
PROBE
2)
Measurement terminals class II L/N against PROBE 2
Measurement range
0.25 ... 20.00 mA
Display range
0.25 ... 20.00 mA
Resolution 0.01
mA
Accuracy
+/- (5% rdg. + 3 digits)
Operational error
± 12%
Limit value
Adjustable 0.25 ... 20.00 mA
Standard limit values:
3.50 mA (class I)
0.50 mA (class II)
Open-circuit test voltage
Approx. 45 V / 50 Hz at 230 V, SELV (floating
output)
Short-circuit test current
Approx. 300 mA
Protection against external
voltage between test terminals Internal fuse 315 mA
Protective Conductor Current TRMS (I
PE(
)
)
Measurement range
0.25 ... 20.00 mA
Display range
0.25 ... 20.00 mA
Resolution
0.01 mA
Accuracy
+/- (5% rdg. + 3 digits)
Operational error
± 12%
Load influence
+/- 0.01 mA / A
Limit value
Adjustable 0.25 ... 20.00 mA, standard limit
value 3.50 mA
Frequency range
40 Hz ... 100 kHz (characteristics according to
DIN EN 61010-1 Annex A, fig. A.1 and VDE 0411
Part
1)
Measurement method
Differential
Mains cord polarity exchange
Automatic by pressing the POL menu key
(GT-800
only)
Mains on switch test (DUT)
Yes, the test done by measuring L current, limit
value 25 mA
Over-range limit
Yes, the measurement will be automatically
interrupted 2 s after exceeding measurement
range
Touch Current TRMS (I
T
)
Measurement range
0.10 ... 2.00 mA
Display range
0.02 ... 2.00 mA
Resolution
0.01 mA
Accuracy
+/- (5% rdg. + 3 digits)
Operational error
± 13%
Limit value
Adjustable 0.02 ... 2.00 mA, standard value
0.50
mA