Accessing the serviceable components of the NObreath®
Using the screwdriver provided with the NObreath®, turn the quarter-turn lock anti-clockwise to
open and slide and lift off the back cover.
Breath drying cartridge replacement
Grip the breath drying cartridge as shown below and tug lightly until you hear a click. Then, simply
lift the cartridge out and replace by sliding a new one into place.
NO scrubber replacement
As shown below, press down on clip on the left-hand side of the monitor and slide out the NO
scrubber then lift to remove and replace with a new one.