16
Ou
r fa
m
ily
, in
no
va
tin
g h
ea
lth
, fo
r y
ou
rs.
the NObreath® has been designed to function in environments containing less than 350ppb NO.
Press
to return the previous screen or
to return to the main screen.
Screen brightness
To adjust the brightness of the display, once in the settings menu, press
to access the
second page where you will find the brightness settings
. Use the arrows
or
to
increase/decrease the brightness of the display. Press
to return the previous screen or
to return to the main screen.
Bluetooth
To turn Bluetooth ON/OFF, once in the settings menu, press
to access the second page and
then press
or
to toggle Bluetooth ON or OFF. If Bluetooth is enabled, the Bluetooth icon
will be visible at the top left-hand of the monitor, next to the battery symbol
. To pair
another Bluetooth enabled device with the NObreath®, select the Bluetooth PIN icon
to
show the pairing PIN onscreen. Press
to return the previous screen or
to return to the
main screen.
Maintenance
•
The NObreath® should be calibrated at least annually or the NO sensor should be replaced.
•
The breath drying cartridge and NO scrubber should be replaced annually.
NOTE:
The unit is designed to provide maintenance reminders when re-calibration is
required. You will be prompted to service the unit when necessary.
•
The NO sensor and pump should be replaced every 5 years
•
The maintenance reminder screen shall be displayed at switch-on if the service
date is due in ≤ 30 days & displayed every day until it has been reset by performing a
calibration or sensor change.