28 User Manual - Beamex
®
CENTRiCAL and Supporting Modules
– Part 5, Pressure Output Modules
S
PECIFICATIONS
G
ENERAL
Feature
SI Unit
Imperial
Width of module
528 mm
20.78"
Height of module
223 mm
8.78"
Total weight of module
3.25 kg
7.2 lbs
Temperature range, operating.
Storage
10 to 60 °C
-20 to 60°C
50 to 150 °F
-4 to 140 °F
Sensitivity better than
(depends on user)
1 mbar
0.015 psi
Maximum pressure for the delivered
pressure hose
20 barg
290 psi g
P
RESSURE
A
DJUSTER
Feature
SI Unit
Imperial
Medium
Clean dry air or other inert,
non-toxic, non-corrosive gas
If you are using Nitrogen, see Warnings in part Appendix of this
manual.
Maximum supply pressure
10 bar
145 psi
Minimum supply pressure
Set pressure
+ 0.5 bar
Set pressure
+ 7¼ psi
Regulating pressure range
0.1 to 8 bar
1.45 to 116 psi
Air consumption (discharged to the atmosphere)
● Bleed flow, max (ANR)
1
with
10 bar (14.5 psi) supply
3.5 l/min
0.124 ft
3
/min
● Exhaust flow, max (ANR) at
maximum set pressure
0.9 l/min
0.032 ft
3
/min
1
ANR = "Atmosphere Normale de Reference",
in English: "standard reference atmospheric conditions".
Summary of Contents for CENTRICAL
Page 2: ......
Page 7: ...Things discussed in this part About this manual Unpacking CENTRiCAL Part 1 INTRODUCTION ...
Page 15: ...CENTRiCAL Dimensions 9 The dimensions for the Corner CM and CF models are also the same ...
Page 28: ...22 User Manual Beamex CENTRiCAL Supporting Modules Part 4 Pressure Measurement Modules ...
Page 29: ...Things discussed in this part Using Pressure Output Modules Part 5 PRESSURE OUTPUT MODULES ...
Page 58: ...52 User Manual Beamex CENTRiCAL and Supporting Modules Appendix ...
Page 59: ...Notes 53 Notes ...
Page 60: ...54 Notes ...
Page 61: ...Notes 55 ...