To install or remove a module from your CENTRiCAL make sure both CENTRi-
CAL and MC6 Workstation are shut off. Disconnect the mains cables from the
wall outlet.
Warning!
Disconnect the mains cables and the pressure supply to
your CENTRiCAL before you install or uninstall any mod-
ules.
R
EMOVING A
P
RESSURE
M
EASUREMENT
M
ODULE
The Pressure Measurement Modules are connected only to the MC6 Workstation
host module.
To disconnect the 20 pin connector from the Pressure Measurement Module:
Unscrew the screws that holds the Pressure Measurement Module in the
CENTRiCAL
’s Instrument Panel (located in the corners of the front panel)
Remove the module from the Instrument Panel. Note that the communi-
cation cable is long enough to allow pulling out the Pressure Measure-
ment Module out without problems.
1. Pull the locking hooks away from the cable connector
2. Pull the cable connector off from the Pressure Measurement Module.
MC6 Workstation Host Module can be used without the removed module, if
necessary. At startup MC6 Workstation automatically identifies all the pressure
modules currently connected to it.
Disconnecting a Pressure Module.
1
1
2
2
0
1
3
4
5
6
7
8
9
ID number
Summary of Contents for CENTRICAL
Page 2: ......
Page 7: ...Things discussed in this part About this manual Unpacking CENTRiCAL Part 1 INTRODUCTION ...
Page 15: ...CENTRiCAL Dimensions 9 The dimensions for the Corner CM and CF models are also the same ...
Page 28: ...22 User Manual Beamex CENTRiCAL Supporting Modules Part 4 Pressure Measurement Modules ...
Page 29: ...Things discussed in this part Using Pressure Output Modules Part 5 PRESSURE OUTPUT MODULES ...
Page 58: ...52 User Manual Beamex CENTRiCAL and Supporting Modules Appendix ...
Page 59: ...Notes 53 Notes ...
Page 60: ...54 Notes ...
Page 61: ...Notes 55 ...