BDDF00122 Issue 3
31/36
Alaris™ Gateway Workstation v1.3.x
Product Specifications
Product Specifications
Electrical
Protection Against Electrical Shock:
Class I
Supply Voltage:
115-230V, ~50-60Hz
Rating:
460VA (Maximum)
Fuses:
2 x 4AH Time-Lag T
Consult Qualified Service Personnel for replacement of fuses .
AC Outlets:
MDI Tile: 115-230V, ~50-60Hz, 60VA
Protection against fluid ingress:
IP22 - Protection against fingers or other object not greater than 80mm in length and 12mm in diameter . Protection against water
droplets deflected up to 15° from vertical . Protected against access to hazardous parts with finger .
Battery
Type:
Nickel Metal Hydride
Charge time:
With no Pumps attached: Up to 8 hours to 95% charge time based on battery life cycle .
With Pumps attached: Up to 16 hours to 95% charge time based on battery life cycle and an 11 Pump load .
Operating time:
60 minutes
Environmental
Operating
Transport and Storage
Temperature:
+5°C - +40°C
-20°C - +50°C
Humidity:
20% - 90%
15% - 95%
Atmospheric Pressure:
700 - 1060hPa
500 - 1060hPa
Classification
Continuous Operation
Regulatory Compliance
Complies with IEC/EN60601-1, IEC/EN60601-1-2 and Radio Equipment Directive 2014/53/EU (where applicable)
Potential Equalisation Conductor
The function of the Potential Equalisation Connector (Conductor) is to provide a direct connection between the Workstation and the
potential equalisation busbar of the electrical installation . To use the Potential Equalisation Connector, connect the Potential Equalisation
Connector on the Workstation to the potential equalisation busbar of the electrical installation .
Patents
Covered under the following Patents: EP 1502612 (DE, CH, FR, GB, ES & IT), CA 2614658 . Design Patents: DE 49910883; FR 997137; and JP
1117998