Frauenfeld, Switzerland
Number of samples until the correct value will be output:
-
In PRECISION mode HIGH, the distance has to be stable for 2 + 0 samples for the output to show the
current value.
-
In precision mode very high, the distance has to be stable for 2 + 16 samples for the output to show
the current value.
Example
Calculation of the response time with measuring frequency 500 Hz, PRECISION = High
1 / 500 Hz =
0.002 s
Median = 7 / 2 (Formula: Samples / 2) =
4
Average =
16
Response time = 0.002 * (4 + 16) = 0.04 s = 40 ms