EMD-RMTR.. User’s Manual
15
51
Anlg Range
Set count
5
0.0000
~
3.9999
52
Meter Fact
Set count
5
0.0000
~
5.9999
53
MeterCode 1
Factory set
6
Finished Y /M
54
MeterCode 2
Factory set
6
Product Serial No
Parameters of converters can decide the running status, process and output ways as well as state of output.
Correct op on and se ng of parameters can keep the converters running op mally and get higher accuracies
of output bother in display and in measurement.
There are 6 grades of passwords for se ng parameters func on. Grades 1 to grade 5 of passwords are for
users and grade 6 of password is for manufacturer. Users can reset their passwords of grades 1 to 4 in grade 5.
Users can check converters parameters in any grade of password. However, if users want to change
parameters pf converters, deferent grade of parameters have to be used by the users.
Grade 1 of password (set by manufacturer as 00521): users can only read parameter.
Grade 2 of password (set by manufacturer as 03210): users can change 1~24 parameters.
Grade 3 of password (set by manufacturer as 06108): users can change 1~25parameters.
Grade 4 of password (set by manufacturer as 07206): users can change 1~38parameters.
Grade 5 of password (Fixed): users can change 1~52 parameters.
Password Grade 5 can be set by skilled users. Grade 4 is mainly used for rese ng total volume in password.
Grades 1~3 can be set by any one who can be chosen by users.
6.2 Details Parameters
6.2.1 Language
There are 2 languages for –RM series converter opera on. They can be set by users according to the users
needs.
6.2.2 Comm Addres
It means this instrument’s address when communicates with many, and has 01~99, holding the 0.
6. 3.3 Baud Rate: 600, 1200, 2400, 4800, 9600, 19200
6.3.4 Snsr Size: 3mm to 3000mm
6.3.5 Flow unit: (L/s
、
L/m
、
L/h
、
m3/s
、
m3/m
、
m3/h)
6.2.6 Flow Range
(Change this value will affect other parameters or output)
Flow range means upper limit value, and lower limit value is set “0” automa cally.
Pulse output will not affect.