2-8
White pin into the REF Electrode Socket
Electrocapillary measurements can be done using the 303A, but only with manual
control; that is, toggle the drop function switch to the DME position just before
the start of the experiment and toggle to either SMDE or HMDE when the scan is
complete.
Accessories
This port allows the BAS 100B/W to sense and control the accessories and
special function modules available to the BAS 100B/W. The accessories that use
this custom port are the Low-Current Module, the Power Module/Potentiostat, the
Rotating Disk Electrode and the AC Impedance Module. More details are
included in each of the accessories operations manual.
Plotter/Printer Connection
These ports are now obsolete as plotting and printing is controlled by the
Windows software. Plotters and printers should therefore be connected to the PC.
Data Link
This port is used to communicate with the external PC via the cable and one of
the serial communication ports (COM ports) on the PC via a standard RS-232C
cable. This connection is shown in Figure 2-7.
Figure 2-7. Attaching remote control (DATA LINK) cable to rear panel connnector.